...B1.2 Description of Parameters
Main Menu
Test
C >
#
Information.
If the sensor is empty or disconnected, the alarms 'MtSnsr' and Coil will be displayed as appropriate.
*
#
The maximum which can be entered must not exceed 21000. The value entered may be displayed with a small error in the
decimal digits e.g. 1.900 may be displayed as 1.899. This is a display characteristic and the value 1.900 will be used by the
MagMaster.
Sub-menus
System Test: Exercises all outputs (as programmed) from manually entered flowrate.
C
If set to '1', transmitter is in 'Test Mode'
Test Mode
1 >
Displays current flowrate: if in 'Test Mode', then any value may be
Test Flow
2 >
entered manually
Flowrate as a percentage.
Test %
3 >
Output Frequency.
Test Hz
4 >
Output Current.
Test mA
5 >
Flow velocity in sensor (for the flowrate above).
Test Vel
6 >
Shows currently active alarms, sequentially
Test Alm
7 >
('Clr' indicates no alarms active).
Live flow velocity (uncorrected for sensor calibration).
Test Txv
8 >
Enter Q (CR) to return to Main Menu or (CR) to cycle around this menu.
Test Quit
Q
Mode' self-cancels after 30 minutes, if no further entries are made.
'Test
END OF PARAMETERS
Remarks
*
...APPENDIX B
Calculated from
'Test Flow 2' above
23