TITLE: Procedure, 5270 Configuration for 4207D
Instrument Model: Model 4207D Compressive Strength Tester
Define the signal I/O addresses as shown. Note that the address for the controller (Load – Programmed) is 1:2f
to acquire the programmed load, the address for the peak value (Load – Maximum) from the load cell is 2:133f.
The address for controller program downloads (Load) is 1.
Verify that the signal I/O address assignments agree with the following: