Specification; Standard Specifications - YOKOGAWA OR8EFG User Manual

Kcl filling type orp sensor
Table of Contents

Advertisement

1.

Specification

The Model OR8EFG KCl filling type ORP Sensor permits stable ORP measurement even for
solutions having comparatively severe properties.
This sensor can be mounted on either an PH8HF flow-through holder or an PH8HS submersion
holder, or its can be used alone suspended in the solution (a maximum depth of 3 meters).
1.1
Standard Specifications
Measurement:
Measurement principle: Metallic electrode method
Measuring range:
Installation:
Note: I f any of the following solutions are measured, install the sensor either in a flow-through or submersion holder.
• When the solution temperature exceeds 80°C.
• When the pH of the solution is 2 or less or 12 or greater.
• When a strong acid solution is to be measured
(e.g., aqua regia, chromic acid, hypochlorous acid or perchloric acid, etc.).
• When the solution contains corrosive gases
(e.g., ammonia, chlorine, hydrogen sulfide, etc).
• When the solution contains a small percentage of organic solvent or oil.
Solution temperature: -5 to 105°C
Table 1. Process Temperature Range
Holder Type
Guide-pipe (PH8HG)
Submersion (PH8HS)
Flow-through
(PH8HF) (
*1)
Suspension (HH350G)
Float (PB350G, PB360G)
Note: PVC: Rigid Polyvinyl, PP: Polypropylene, SS: Stainless Steel.
*1:
For flow-through types, refer also to the solution temperature and pressure graph
(in notes following Model and Suffix code table for flow-through type holders).
*2:
Solutions with normal pH ranges of 3 to 14 are recommended for stainless steel (316 SS).
Oxidation-reduction potential of a solution
-1500 to 1500 mV
Mounting in PH8HS submersion holder
Mounting in PH8HG guide-pipe holder
Mounting in PH8HF flow-through holder
when mounted in PH8HG guide-pipe holder: -5 to 80°C
when mounted in PH8HS, PH8HF holder, see Table 1.
Holder
Material
PVC
PP
PP
SS (
*2)
PP
SS (
*2)
SS (*2)
None, Provided
PP, SS (*2)
< 1. Specification >
Solution
Cleaner
pH Range
None
2 to 12
None
Provided
None
Provided
None
0 to 14
Provided
None
Provided
None
Solution Temperature
(°C)
-5 to 50
-5 to 80
-5 to 100
-5 to 80
-5 to 100
-5 to 80
-5 to 80
-5 to 80
-5 to 105
-5 to 80
-5 to 80
-5 to 50
IM 12C07J01-01E