8.2 Muting the Alarms

Press the Mute button to mute alerts and alarms.
Pressing the mute button will only silence the alarm for 24 hours if the alarm condition
remains active during that time. If the alarm clears and returns, the alarm will sound. If a new
alarm occurs the alarm will sound.

8.3 Alarm and Notification Types

All alarms generated are defined as low priority alarms.
The following table contains general descriptions and solutions for the categories of alarms.
Error
Type
Code
Running
LXXXX
Alarms
Shut Down
HXXXX
Alarms
Notifications
LXXXX
60126903-A
Solution
1. Check if flowmeter is set between 1.0-5.0
L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages (kinked
cannula, blocked vents, etc.)
4. Restart unit using stop/start button
5. Ensure device is in specified operating
environment
If alarm is still present, contact provider
1. Check if flowmeter is set between 1.0-5.0
L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages (kinked
cannula, blocked vents, etc.)
4. Restart unit using stop/start button
5. Ensure device is in specified operating
environment
If alarm is still present, contact provider
Contact provider for service
Invacare® Platinum® 5NXG Oxygen Concentrator
Indicator
Symbol
Light Color
White
with
yellow
dot
White
with red
dot
White
with blue
dot
Sound
Indication
Beeping
Constant
Audible
Alarm
N/A
36