Measurement uncertainty is included in the device specification. All conditions at STPD (standard
temperature and pressure, dry).

9.2 Electromagnetic Compatibility (EMC)

The Platinum® 5NXG Concentrator requires special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in these accompanying documents.
Guidance and manufacturer´s declaration—electromagnetic emission
The Platinum® 5NXG Concentrator is intended for use in the electromagnetic environment specified below. The
customer or the user of the Platinum® 5NXG Concentrator should assure that it is used in such an environment.
Emissions
Compliance
test
RF emissions
Group I
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic
emissions IEC
Class A
61000-3-2
Voltage
fluctuations
Complies
IEC 61000-3-
3
43
Humidifier Bottle/water
Flowmeter
Control Panel/Display/Buttons*
Shroud/Base
Power Cord
Circuit Breaker Switch
Casters/Pegs
* Temperatures of these parts under a single fault condition of
blocked air inlets can reach 144° F (62° C).
Electromagnetic environment - guidance
The Platinum® 5NXG Concentrator uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The Platinum® 5NXG Concentrator is suitable for use in all establishments,
including domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.