Summary of Contents for Invacare Platinum Mobile POC1-100 Series
Page 1
Invacare® Platinum® Mobile POC1–100 Series en Oxygen Concentrator Service Manual DEALER: Keep this manual. The procedures in this manual MUST be performed by a qualified technician.
Page 2
All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
Invacare® Platinum® Mobile Refer to the following table for the meaning of symbols 1 General marked on the equipment and/or packaging. 1.1 Symbols General Warning Sign Signal words are used in this manual and apply to hazards The background color inside the or unsafe practices which could result in personal injury triangle is yellow on product labels.
Page 5
General DO NOT Reuse (Single Patient Use Only) Class II Equipment This symbol refers to the cannula which is for single patient use only. Power On/Off Latex Free Data Output Terminal Indoor Use ONLY This symbol is included on the AC DO NOT dispose of in household waste power adapter.
Invacare® Platinum® Mobile On/Off Button Transport and Storage Humidity Press for one second to turn concentrator on or off. Flow Selection Button This product complies with Directive 93/42/EEC concerning medical devices. Press to adjust the pulsed oxygen flow to your prescribed setting.
General 1.2.2 Input/Output Connections Battery Gauge Indicates the charge status of the battery(s). Refer to Reading the Battery Gauge for Installed and Uninstalled Battery(s) in Setup for more details. External Power Indicator The orange indicator light next to the plug symbol illuminates when an external power source is connected to the concentrator.
Invacare® Platinum® Mobile 1.2.3 Filters Air enters the concentrator through an air intake filter located on the carry bag. This filter prevents hair and other large particles in the air from entering the unit. Before you operate the concentrator, make sure this filter is clean and dry, and the concentrator is properly installed in the carry bag.
Safety 2 Safety DANGER! Risk of Death, Injury Or Damage From Fire Textiles, oil or petroleum substances, grease, 2.1 General Guidelines greasy substances and other combustibles are easily ignited and burn with great intensity in In order to ensure the safe installation, assembly and oxygen enriched air and when in contact with operation of the concentrator these instructions MUST be oxygen under pressure.
Page 10
Any other accessories have oxygen under pressure. To avoid fire, death, not been tested by Invacare and are not injury or damage: recommended for use with Invacare products. – Avoid creation of any spark near oxygen –...
Page 11
Safety WARNING! WARNING! Risk of Death, Injury or Damage Risk of Injury or Damage To prevent injury or damage from cord misuse: To avoid injury or damage from airborne – DO NOT move or relocate concentrator by pollutants and/or fumes and for optimal pulling on the cord.
Invacare® Platinum® Mobile 2.2 Radio Frequency Interference Medical Electrical Equipment needs to be installed and used according to the EMC information in this manual. This equipment has been tested and found to comply with EMC limits specified by IEC/EN 60601-1-2. These limits...
Setup 3.2 Changing the Date and Time 3 Setup This procedure describes the process to change the 3.1 Viewing the Date and Time date and time. This procedure displays the date and time. The concentrator must be Off and connected to external power to perform this procedure.
Page 14
Invacare® Platinum® Mobile 4. Press the flow selection button D to increase the selected value by one until the maximum is reached and then rolls over to the lowest value. 5. Press the display backlight button to decrease the selected value by one until the minimum is reached and 1.
Setup 8. When the time and date have been adjusted as desired, press the power button to highlight Set. 9. Press the flow selection button to save the date and time and to exit the time and date screen. 10. Highlight Cancel and press the flow selection button to cancel setting the time and date and to exit the time and date screen.
Page 16
Invacare® Platinum® Mobile 3. Press and release the On/Off button A to enter the Language Change Screen. The proposed language selection is highlighted. 4. Press the flow selection button D to move to the next language selection. Press the display backlight button C to move to the previous language selection.
Page 17
Setup 6. Press the flow selection button to set the display language to the highlighted language and exit the Language Change Screen. If you do not want to change the display language, press the On/Off power button to highlight “CANCEL”. Then press the flow selection button to exit the Language Change Screen.
4.2 Service and Replacement Parts Improper maintenance and servicing may lead to injury or damage. To avoid injury or damage: Refer to the parts catalog at www.invacare.com for – Use Invacare service parts only. service and replacement part information. – Only professionals of the health care field or persons fully conversant with this process 4.3 Service Life...
Page 19
Service 5. Insert T-20 driver into opening of oxygen outlet port I 10. Reinstall barbed cap and hand tighten with T-20 driver and turn counter-clockwise until barbed cap J comes until the barbed cap fits firmly against the lower half free from the lower half of the port.
The spring I may come out of the DC power cable when the fuse is removed. Only use fuses provided by Invacare. 3. Ensure the spring is installed into the DC power cable. 4. Seat the fuse into the tip G.
Service 4.6 Replacing the Battery Pull Ring 3. Remove the existing pull ring from the battery pack. 1. Lift the rounded end of the existing pull ring. 4. Press the new pull ring down to engage into the mount. 2. Slide the existing pull ring out of its mount. 1187375-E...
The POC1-CONNECT is not available in all regions. The POC1-CONNECT dongle is only compatible with Invacare Platinum Mobile oxygen concentrators beginning with serial number 17JF029716. 1. Remove the Platinum Mobile oxygen concentrator from the carry bag. Refer to the user manual.
Troubleshooting 5.1.1 Viewing the Current Alarm Code and Software 5 Troubleshooting Version Number 5.1 Alarm Codes This procedure displays the alarm code of any active alarm condition with the software version number. Basic troubleshooting and alarm condition information is contained in the user manual. The concentrator may be Off and connected to external power or turned On to perform this The Platinum Mobile concentrator has dozens of possible...
Invacare® Platinum® Mobile Sample Alarm Code Display Screens Software Version Bootloader Version 5.1.2 Viewing the Alarm Code History The initial “0x” portion of the alarm code is not This procedure displays the 20 most recent alarm shown in the alarm code history screens.
Page 25
Warm Up") 2. Restriction after through his/her the conserver nose. valve. 2. Check the outlet filter/Replace filter. 3. If alarm persists, contact Invacare for service. 0x4202 Breath LPA_breathRateTooHigh The display shows: 1. The user breath 1. Make sure breath Rate Too "Breath Rate High...
Page 26
Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name 0x7207 LPA_lowBattery30MinLeft The display shows: 1. The concentrator 1. Supply unit with Battery "Low Battery Charge is running off external power (~30 Unit" battery power and...
Page 27
To Charge to Cool" external power to cool. (Running) (AC/DC) and 2. If alarm persists, at least one contact Invacare of the battery for service. temperatures is greater than or equal to 49° C. Battery 0xb20c LPA_batteryTooCold- The display shows: 1.
Page 28
Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name Battery 0x6206 LPA_batteryCold The display shows: 1. The concentrator 1. Remove the cold Cold "Battery Cold Allow to is running off battery and allow (Running) Warm"...
Page 29
3. Change outlet power-on state. filter. a. Unit running 4. If alarm persists, without cannula contact Invacare connected -or- for service. b. Inlet airflow blockage -or- c. Outlet filter perforated -or- d. Sieve bed has been...
Page 30
Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name Prolonged 0x9209 LPA_prolongedUnit- The display shows: 1. Concentrator is 1. Ensure cannula is producing less Unit Warm WarmUp "Warming Up" connected while than 87% O after unit is operating.
Page 31
Name Oxygen 0x8208 LPA_oxygenSensorFailure The display shows: 1. Oxygen sensor 1. If alarm persists, Sensor is damaged or "Sensor Fault See contact Invacare Failure disconnected. Manual" for service. 0x1100 Depleted SDA_depletedBattery The display shows: 1. The concentrator 1. Supply unit with Battery "Shut Down Battery...
Page 32
Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name Battery 0x1102 SDA_batteryTooCold The display shows: 1. The concentrator 1. Remove the cold Too Cold "Shut Down Battery is running off battery and allow (shut Too Cold"...
Page 33
15 minutes after filter screen. the power-on 3. Change outlet state. filter. a. Unit running 4. If alarm persists, without cannula contact Invacare connected -or- for service. b. Inlet airflow blockage -or- c. Outlet filter perforated -or- d. Sieve bed has been...
Page 34
Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name Stuck 0x1107 SDA_stuckButton The display shows: 1. One of the 1. Make sure no Button "Shut Down Stuck buttons on the control panel Button"...
Page 35
Fault" Blockage may cannula. be internal or 2. Replace the external. patient outlet filter. 3. If alarm persists, contact Invacare for service. 0x1121 Internal lowPressureAlarm The display shows: 1. Internal leakage of 1. Clean air intake Fault "Shut Down Internal pneumatic circuit.
Page 36
Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name Internal 0x1123 compressorNotRun- The display shows: 1. Compressor or 1. If alarm persists, ningAlarm compressor's Fault "Shut Down Internal contact Invacare wire harness is Fault"...
Maintenance 6.3 Cleaning the Air Intake Filter Screen 6 Maintenance CAUTION! 6.1 Wear and Tear Risk of Damage To avoid damage from clogging: Refer to the User Manual for the list of normal wear and – Replace the carry bag if the air intake filter tear items and components for this product.
Invacare® Platinum® Mobile 6.4 Cleaning and Disinfection Between Patients 3. Disinfect the surfaces of the concentrator and accessories using Clorox® 4 in One Disinfectant & Cleaning and disinfection must be performed between Sanitizer or substantially equivalent product. patients and as necessary.
Clean Air Intake Filter Screen. Inspect Cabinet of Concentrator and AC Adapter for Damage. Do not use if damage found. Return to Invacare for repair. Inspect Carry Bag Strap(s) and hardware for Damage. Replace strap or carry bag if damage is found.
Page 40
Invacare® Platinum® Mobile EVERY 4,380 HOURS, EVERY 3 YEARS, AND BETWEEN PATIENTS, WHICHEVER COMES FIRST Check Oxygen Purity Perform Functional Test DURING PREVENTATIVE MAINTENANCE SCHEDULE AND BETWEEN PATIENTS Check Intake Filter Screen for damage. Replace carry bag if damage is found.
87% or above after a five minute warm-up period. 8. If oxygen purity level is 87% or less, check the air intake filter screen and clean if necessary. If still getting an oxygen purity level less than 87%, contact Invacare for service. 1187375-E...
Invacare® Platinum® Mobile 6.7 Function Test 11. Breathe with the cannula to ensure the unit pulses at each pulse flow setting. Confirm that the pulse 1. Install a battery in the unit and connect to AC power icon is shown on the display screen with each oxygen source using the AC power adapter.
After Use 7 After Use 7.1 Disposal Recycle DO NOT dispose of in household waste This product has been supplied from an environmentally aware manufacturer who complies with the Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EC. This product may contain substances that could be harmful to the environment if disposed of in places (landfills) that are not appropriate according to legislation.
Page 47
Ireland • Tel: (353) 1 810 7084 • Fax: (353) 1 810 7085 • ireland@invacare.com • www.invacare.ie Italia Invacare Mecc San s.r.l., Via dei Pini 62, I-36016 Thiene (VI) • Tel: (39) 0445 38 00 59 • Fax: (39) 0445 38 00 34 • italia@invacare.com • www.invacare.it Nederland Invacare BV, Galvanistraat 14-3, NL-6716 AE Ede •...
Page 48
Invacare Corporation Australia Manufacturer One Invacare Way Invacare Australia Pty Ltd Invacare Corporation Elyria, Ohio USA 1 Lenton Place 2101 E. Lake Mary Blvd. 44035 North Rocks NSW 2151, Sanford, FL USA 32773 440–329–6000 Australia 407–321–5630 800–333–6900 1800–460–460 (tel) Technical Services 1800–814–367 (fax)