Stacking Concentrators - Invacare Platinum P5NXG1 User Manual

Stationary oxygen concentrator
Table of Contents

Advertisement

Invacare® Platinum® 5NXG Oxygen Concentrator

4.4 Stacking Concentrators

WARNING!
Risk of Injury or Damage
This feature must only be used in a commercial environment for storage purposes only. As such, this
feature has been evaluated only under the following conditions:
In low traffic areas and on flat, smooth, unbroken, hard ground-level surfaces
l
Subject to no external disturbances, or being bumped, moved, shaken, or subject to external force
l
Stacked no more than two units high, including the lowest level units, as shown in the
l
instructions for use
Stacked only with the same units and not mixed with other models
l
Properly nested as shown in the instructions for use
l
Stacked units in non-operating state/no power applied
l
Not moved, transported, or repositioned when stacked
l
Any other use of stacking feature is considered misuse of the product and may cause instability,
damage, or injury.
Units shall not be operated while in a stacked position.
These directions do not alter, replace, modify, or displace safe work practices, which include:
Keeping a safe distance between people and the stacked concentrators
l
The need for safety equipment, such as reinforced-toe shoes
l
Periodic inspection to ensure proper storage conditions are present
l
Storage in a place not subject to conditions that would make stacking unsafe
l
This procedure is for providers and qualified service personnel only.
1. Position two units side-by-side with the front of the each unit
facing toward you. The units shall be positioned so that a small
gap exists between units.
15
Fig. 4-1 Position units side-by-side
60126903-A

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Platinum p5nxgc1Platinum 5nxg

Table of Contents