Invacare® Platinum® 5NXG Oxygen Concentrator

9 Technical Data

9.1 Specifications

Electrical Requirements:
Rated Current Input:
Sound pressure level (typical)
Sound pressure level
(per ISO 80601-2-69):
Sound power level
(per ISO 80601-2-69):
Altitude:
Oxygen Output Concentration Levels:
Maximum Outlet Pressure:
Flow Range:
Potential Obstruction Alert:
Power Consumption:
Pressure Relief Mechanism Operational at:
Change in maximum recommended flow
when back pressure of 7kPa is applied:
Filters:
Safety System:
41
120 VAC + 10, –15% (132 VAC/102 VAC), 60 Hz
3.0 A
< 44 dB(A)
Max 60 dB(A)
Max 60 dB
Up to 13,000 ft (3962 m) above sea level.
While filling a HomeFill System: Up to 6,000 ft (1828 m) above
sea level.
Atmospheric Pressure Range: 101.3 kPa – 62.0 kPa
87% to 96% at all patient flow rate settings
Concentration levels achieved after initial start-up period
(between 5-30 minutes)
5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
1 to 5 L/min (maximum). For flowrates less than 1 L/min, we
recommend the use of the Invacare PreciseRx Pediatric
Flowmeter (IRCPF16).
0 L/min to 0.5 L/min
The concentrator detects a condition that may indicate a
potential obstruction of the output oxygen. Rapid audible
beeping alert (this alert is deactivated when accessories are
connected) may be associated with flow setting of 0.5 L/min or
less.
245 W at ≤ 2 L/min
260 W at 2-3 L/min
285 W at > 3 L/min
35 psi to 45 psi (242.3 kPa to 310.3 kPa )
0.7 L/min (Back Pressure of 7 kPa applied)
Cabinet, Patient Outlet and Compressor Inlet
Current overload or line surge shutdown.
High Temperature Alarm
High Pressure Alarm
60126903-A