Invacare® Platinum® 5NXG Oxygen Concentrator
Alarm
High Pressure
High Presure Diagnostics
Low Pressure Diagnostics
High Temperature Shut
Down
Valve Coil Failure
Fan Not Connected
Fan Speed Out of Range
39
Error
Solution
Code
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages
H3020
(kinked cannula, blocked vents,
etc.)
4. If alarm is still present, contact
provider
Contact provider/Refer to Service
H302x
Manual
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages
H304x
(kinked cannula, blocked vents,
etc.)
4. If alarm is still present, contact
provider
1. Inspect for blocked vents
2. Ensure device is in specified
H4000
operating environment
3. If alarm is still present, contact
provider
1. Restart unit using stop/start
button
H5000
2. If alarm is still present, contact
provider
1. Restart unit using stop/start
button
H6000
2. If alarm is still present, contact
provider
1. Restart unit using stop/start
button
H6100
2. If alarm is still present, contact
provider
Indicator
Symbol
Light Color
Indication
White
with red
dot
White
with red
dot
White
with red
dot
White
with red
dot
White
with red
dot
White
with red
dot
White
with red
dot
Sound
Constant
Audible
Alarm
Constant
Audible
Alarm
Constant
Audible
Alarm
Constant
Audible
Alarm
Constant
Audible
Alarm
Constant
Audible
Alarm
Constant
Audible
Alarm
60126903-A

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Platinum p5nxgc1Platinum 5nxg

Table of Contents