8.4 Error Codes

The following table contains detailed descriptions and solutions for the error codes and alarms.
Alarm
Running Alarms
Potential
Obstruction Alert
Oxygen Sensor
Heater Open Failure
Oxygen Sensor
Heater Short Failure
Oxygen Sensor Out
of Range High
Failure
Oxygen Sensor Out
of Range Low Failure
Low Oxygen
Concentration
Low Oxygen
Concentration
Diagnostics
37
Error
Solution
Code
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or
L1000
blockages (kinked cannula,
blocked vents, etc.)
4. If alarm is still present,
contact provider
L2000
L2010
1. Restart unit using
stop/start button
2. If alarm is still present,
L2020
contact provider
L2030
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or
L2040
blockages (kinked cannula,
blocked vents, etc.)
4. If alarm is still present,
contact provider
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages
L204x
(kinked cannula, blocked
vents, etc.)
4. If alarm is still present,
contact provider
Indicator
Symbol
Light Color
White
Single audible beep with
with
yellow
dot
White
Single audible beep with
with
yellow
dot
White
Single audible beep with
with
yellow
dot
White
Single audible beep with
with
yellow
dot
Sound Indication
less than 2 second
interval.
less than 20 second
interval.
less than 20 second
interval.
less than 20 second
interval.
60126903-A