Alarm
High Temperature
Alert
Alarm
Shut-Down Alarms
Power Loss 3 times in a row
Low Oxygen Concentration
Shut Down
Low Oxygen Concentration
Shut Down Diagnostics
Pressure Sensor Out of
Range High
Pressure Sensor Out of
Range Low
60126903-A
Error
Solution
Code
1.
Inspect for blocked vents
2.
Ensure device is in specified
L4000
operating environment
3.
If alarm is still present,
contact provider
Error
Solution
Code
1. Restart unit using stop/start
button
H1000
2. If alarm is still present, contact
provider
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages
H2040
(kinked cannula, blocked vents,
etc.)
4. If alarm is still present, contact
provider
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages
H204x
(kinked cannula, blocked vents,
etc.)
4. If alarm is still present, contact
provider
1. Restart unit using stop/start
button
H3000
2. If alarm is still present, contact
provider
1. Restart unit using stop/start
button
H3010
2. If alarm is still present, contact
provider
Invacare® Platinum® 5NXG Oxygen Concentrator
Indicator
Symbol
Light Color
White
Single audible beep with
with
yellow
dot
Indicator
Symbol
Light Color
White
with red
dot
White
with red
dot
White
with red
dot
White
with red
dot
White
with red
dot
Sound Indication
less than 20 second
interval.
Sound
Indication
Constant
Audible
Alarm
Constant
Audible
Alarm
Constant
Audible
Alarm
Constant
Audible
Alarm
Constant
Audible
Alarm
38