2 Description
2.1 Overview
Part
1
2
3
4
5
Manifold
Ceramic Manifold
The manifold (see following figure) is made of aluminum oxide ceramic for efficient
heat transfer and temperature control during IEF. Protrusions along the channel inside
the manifold align the rehydrated IPG strip, keeping it straight and centered when
placed inside the manifold. The manifold has a coating to minimize protein adsorption.
Because some cleaning agents can remove this coating, clean the manifold only with
the Ettan IPGphor Strip Holder Cleaner.
Note:
Manifold Light
The manifold light is made of the polymers PC (polycarbonate) and ABS (acrylonitrile-
butadiene-styrene).
The manifold light's flat design is important for the heat transfer. If exposed to very
high temperatures, the flatness could be affected. Refer to
regarding cleaning temperature.
General information
IPG strips from 7 cm to 24 cm long may be used. The unit may also be used for IPG strip
equilibration prior to second dimension electrophoresis. Forcep wells are located at
the ends of the channels to aid strip removal after electrophoresis.
8
Function
Manifold
Pre-cut electrode wicks
Electrode assembly
Sample cups
Cleaning brush
The manifold is fragile and should be handled with care.
Part
Function
6
Pre-cut paper bridges
7
Sample cup insertion tool
8
Cleaning solution
9
Forceps
10
Spirit level
Manifold Light, on page 23
Ettan IPGphor User Manual 11003460 AE