3.2

Run conditions

Introduction
Program the Ettan IPGphor with the recommended run parameters. Ramping the
voltage slowly while the sample is entering the IPG strip will improve results.
Instructions for optimal ramp, voltages and times, or Vhrs (volt-hours) totals are
attached to each package of Immobiline Dry Strip.
Ceramic manifold
Follow the recommended run conditions found in the Instructions attached to each
Immobiline DryStrip package.
Manifold Light
Follow the recommended run conditions found in the Instructions attached to each
Immobiline DryStrip package except for focusing time.
Focusing time
As a guideline, use the lower number of recommended volt-hours to avoid overfo-
cusing. Normally the focusing results does not improve after this time.
Ettan IPGphor User Manual 11003460 AE
WARNING
Ettan IPGphor is a high voltage instrument that can cause fatal
electrical shock if the safety features are disabled. The safety lid
must be securely closed before starting a protocol.
WARNING
The Ettan IPGphor unit is capable of producing thousands of volts.
Before operating the unit, read and fully understand the Ettan
IPGphor operating instructions and warnings.
CAUTION
Do not use any organic solvents during cleaning.
WARNING
Check that the high voltage is turned off before opening the safety
lid. The HV ON indicator on the front panel should not light.
3 Operation
3.2 Run conditions
21