Three operation modes (cont.)
«ROUTINE» Mode
Select «ROUTINE» Mode to perform automated
duplicate measurements on the preferred arm for
prompt and accurate ofice measurements.
ABI ROUTINE SCREEN
14
Automated triple measurements
In «ROUTINE» Mode, the WatchBP Ofice ABI device
automatically takes three consecutive measurements
at 15 second intervals on default. The readings of
these three measurements are averaged to produce
the routine visit blood pressure measurement.
1
15 sec.
The user can manually set measurement intervals of
*
15, 30, 45 or 60 seconds in ROUTINE Mode. (Please
refer to special functions section page 19 "Setting
measurement intervals")
The number of measurements in ROUTINE mode can
*
be set at two by request. Please contact microlife or
its distributors for further information.
2
3
15 sec.