Three operation modes (cont.)
Automated triple measurements
In «SCREEN» Mode, the WatchBP Ofice ABI
device automatically takes three consecutive
measurements on both arms at ixed one minute*
intervals. The results of these three measurements
are then averaged to conclude the blood pressure
measurement.
1
60 sec.
The measuring intervals are ixed at one minute in
*
SCREEN mode.
The arm with the higher blood pressure value should
*
be taken for ABI measurement or future blood pressure
measurements.
16
2
3
60 sec.
Determine the proper arm and inter-arm
difference
If the difference of blood pressure readings between
two arms is more than 20mmHg for Systolic or
10mmHg for Diastolic at all three measurements,
the device indicates the arm with the higher blood
pressure value by displaying L or R and the "IAD" icon.
The "IAD" icon and readings of the higher arm lash to
indicate that more attention is needed.
IAD
SCREEN
IAD
L