The instrument should be serviced approximately once per year.
O-
RINGS
O-rings on the piston and anvil are essential to the accurate operation of the instrument and should
be checked regularly for signs of wear or damage. The O-rings will wear, especially on the piston,
and hence should be replaced regularly.
To check for O-ring leaks:
1. Tap the Maintenance view, then tap Advanced.
2. Select Allow calibration and tap Save.
3. Tap the Instrument view.
Changing calibrations will affect the performance of the instrument. The following steps
do not change any calibrations. Pressing other buttons on the left side of the display may
change calibrations.
4. Install an empty sample tube.
5. Enter a flow rate of 10 sccm and tap
6. When pressure stabilizes, try to move the tube by hand. The O-rings should be replaced if the
tube moves freely or the inlet or differential pressure vary by more than a few tenths of a cm
H
O.
2
7. Tap
Stop
Flow.
8. Tap the Maintenance view, then tap Advanced.
9. Deselect Allow calibration and tap Save.
A
S
IR
UPPLY
Check that the cylinder has adequate pressure. Replace the cylinder when it is low. See
Gas Cylinder on page 7 - 4
MIC SAS II 5800 Operator Manual
580-42800-01 (Rev - ) — June 2019
Set Flow
.
Rate.

Routine Maintenance

Replace a
7 - 9