Service Mode

Revision C
Installation: Configuring Your ProCare Monitor
Service mode is entered by holding the Cycle button while powering on with the
On/Off button. You can press the Cycle button to advance through the available
service mode pages.
NOTES
Only transducer calibration pages are available until calibration is valid.
Calibration and other service mode setting changes will not be retained
unless the "Save Settings" operation is executed (on the final Service Mode
options page).
After all options pages have been displayed, the display will return to the 1
service mode page (initial calibration page). To save service mode settings the
Service Mode option pages are as follows:
Displayed on Monitor
0 (in "min" window)
1 (in "min" window)
2 (in "min" window)
3 (in "min" window)
4 (in "min" window)
5 (in "min" window)
®
DINAMAP
ProCare Monitor Service Manual
2009381-001
Function
Refer to calibration section for functions.
NIBP Algorithm Type loaded
(Displayed in Systolic display window)
1: Classic DINAMAP NIBP
2: Auscultatory NIBP
(This setting cannot be changed.)
SpO
Type loaded
2
(Displayed in SpO
% display window)
2
0: No SpO2
1: NellCor
2: Masimo
(Warning: Incorrect setting will cause fatal 930 alarm
during operation.)
Temperature loaded
(Displayed in Temperature display window)
0: No Temp
1: Turbo Temp
st
3-11