Display; Cuffs; General; Materials - GE DINAMAP ProCare Service Manual

Table of Contents

Advertisement

Display

Cuffs

Revision C
Maintenance: Cleaning
Any type of solution that contains ammonium chloride, conductive
solutions, wax or wax compounds
Sodium salts
NOTE:Never autoclave or steam clean the monitor, cuffs, or accessories.
To clean the display screen, use a soft, clean cloth dampened with a glass
cleaner. Never spray the glass cleaner directly onto the display, and never use
alcohol or petroleum based products.

General

The cuff must be thoroughly cleaned with the specified detergent before reuse.
The additional use of household bleach as described below provides at least
intermediate-level disinfection.
Apply cuff hose plugs before cleaning.
The following cleansing procedure was repeated 20 times on DURA-CUF
Blood Pressure Cuffs and once on SOFT-CUF
affecting the performance of the cuff.
While this procedure is adequate for cleaning/disinfection, it may not
remove all stains.
Do not immerse hoses.
Do not immerse cuffs without prior application of cuff hose caps.

Materials

Enzymatic detergent such as ENZOL* enzymatic detergent (US) or
Cidezyme* enzymatic detergent (UK)
Distilled water
10% solution of household bleach (5.25% sodium hypochlorite) in distilled
water
Soft cloths and soft-bristled brushes
Spray bottles

Procedure

1.
Prepare the enzymatic detergent according to the manufacturer's
instructions and the 10% bleach solution, in separate spray bottles.
2.
Spray the detergent liberally on device. If the material is dried on, allow the
cuff to sit for 1 minute. For soil on the soft part of the closure or the cuff
itself, wipe the material off with a soft cloth. For persistent contamination on
the soft part of the closure, use a soft-bristled brush to loosen particles.
Rinse with copious amounts of distilled water. Repeat until no visible
contamination remains. For soil on the hook part of the closure, use a soft-
bristled brush to remove the material, and rinse with copious amounts of
distilled water. Repeat until no visible contamination remains.
®
DINAMAP
ProCare Monitor Service Manual
2009381-001
®
Blood Pressure Cuffs without
®