Electromagnetic Compatibility (EMC): Electromagnetic Compatibility (EMC): ILC-1926
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Immunity Test
EN 60601 Test Level
Electrostatic
± 6 kV contact
Discharge (ESD)
± 8 kV air
EN 61000-4-2
Electrical Fast
± 2 kV for power supply
Transient/Burst
lines
EN 61000-4-4
±1 kV for input/output
lines
Surge
± 1 kV differential mode
EN 61000-4-5
± 2 kV common mode
<5% U
for 0.5 cycles
Voltage dips,
<40% U
short
for 5 cycles
interruptions
<70% U
and voltage
variations on
for 25 cycles
power supply
input lines
EN 61000-4-11
<5% U
for 5 s
Power
3 A/m
Frequency (50/
60 Hz) Magnetic
Field
EN 61000-4-8
Revision C
The ILC-1926 is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or user to assure that the ILC-1926
is used in such an environment.
Compliance Level
± 6 kV contact
± 8 kV air
No AC Input
No AC Input
(>95% dip in U
)
<5% U
t
t
U
) for 0.5 cycles
t
(>60% dip in U
)
<40% U
t
t
in U
) for 5 cycles
t
(>30% dip in U
)
<70% U
t
t
in U
) for 25 cycles
t
(>95% dip in U
)
<5% U
t
t
U
) for 5 s
t
3 A/m
NOTE:
U
is the AC mains voltage prior to application of the test level.
t
®
DINAMAP
ProCare Monitor Service Manual
2009381-001
Electromagnetic Environment – Guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
No AC Input
Mains power should be that of a typical
commercial or hospital environment.
(>95% dip in
t
(>60% dip
Mains power should be that of a typical
t
commercial or hospital environment. If the
user of the DINAMAP ILC-1926 requires
(>30% dip
continued operation during power mains
t
interruptions, it is recommended that the
Host monitor be powered from an
uninterruptible power supply or a battery.
(>95% dip in
t
Power frequency magnetic fields should be at
levels characteristics of a typical location in a
typical commercial or hospital environment.
C-9