GE DINAMAP ProCare Service Manual page 75

Table of Contents

Advertisement

Error Code
Source
or Problem
E13
Battery
E21
SPO2
E23
SPO2
E25
SPO2
E61
Temp
E63
Temp
E66
Temp
E80
NIBP
E82
NIBP
E83
NIBP
E84
NIBP
E85
NIBP
E89
NIBP
*Acknowledging an alarm by pressing the Silence button, cancels the alarm.
** Call Technical Support for definitions and instructions.
Revision C
Troubleshooting: Alarm Code Interpretation
Definition
Battery too low to operate the unit
Failed SPO2 sensor
SPO2 Sensor off finger
SPO2 No signal
Temp probe failure
Probe disconnected or wrong type
Temperature probe too hot
Overpressure
Excess air in cuff
NIBP inflation time out
NIBP total time out
NIBP level time out
NIBP no pulse detected
®
DINAMAP
ProCare Monitor Service Manual
2009381-001
Can be
Probable Source
Silenced?
No
Charge battery
External DC source failed
Replace battery
Replace SPO2 sensor
Yes
Reposition SPO2 sensor
Yes
Reposition SPO2 sensor
No
Replace temperature probe
Yes
Check for correct probe
No
Replace probe
Yes
Check for pinched or occluded
internal tubing
Yes
Wrong cuff type
Failed valve on pneumatics
assembly
Yes
Leak in cuff or o-ring in hose
Internal leak in tubing or
pneumatic valve
Pump not turning on
Yes
Pneumatic assembly failure
Yes
Cuff placement on patient
Minor leak in cuff
Pneumatic assembly failure
Yes
Cuff placement on patient
5-5