Calibrating The Meter (Cal); Recalibrating The Internal Calibration Reference (Ical) - FlowLine LI25 Series Instruction Manual

Loop-powered meters
Hide thumbs Also See for LI25 Series:
Table of Contents

Advertisement

LI25 Series Loop-Powered Meters

Calibrating the Meter (Cal)

The meter can be calibrated to display the process in engineering units
by applying the appropriate input signal and following the calibration
procedure.
The use of a calibrated signal source is strongly recommended.
Cal
inpt1
1.
Press the Up arrow button to scroll to the Calibration menu (CAL)
and press Enter/Ack.
2.
The meter displays inpt1. Apply a known signal and press
Enter/Ack. Trend arrows are displayed while accepting the signal.
3.
After the signal is accepted, the meter displays dspl1. Press
Enter/Ack, enter a corresponding display value for the signal input,
and press Enter/Ack to accept.
4.
The meter displays inpt1. Apply a known signal and press
Enter/Ack. Trend arrows are displayed while accepting the signal.
5.
After the signal is accepted, the meter displays dspl2. Press
Enter/Ack, enter a corresponding display value for the signal input,
and press Enter/Ack to accept.

Recalibrating the Internal Calibration Reference (ICal)

The Internal Calibration (ICAL) menu, located in the Advanced features
menu, is used to recalibrate the internal calibration reference.
Recalibration is recommended at least every twelve months. Refer to
Internal Calibration (ICal), page 29 for instructions.
To scale the meter without a signal
source refer to Scaling the Meter
(sCalE), page 20.
Press Enter to Accept Setting
Press Up to Set Digit Value
Press Right to Select Next Digit
Press Menu to Exit at any Time
dspl1
Display
Flashes
Accepting
Input
04. 0 00
Set Display 1
Value
21
Instruction Manual
inpt2
Calibrate
Input 2