FlowLine Symprobe LP75 Owner's Manual

FlowLine Symprobe LP75 Owner's Manual

Rf capacitance level transmitter

Advertisement

Warranty, Service & Repair
To register your product with the manufacturer, fill out the enclosed
warranty card and return it immediately to:
Flowline Inc.
10500 Humbolt Street
Los Alamitos, CA 90720.
If for some reason your product must be returned for factory ser-
vice, contact Flowline Inc. to receive a Material Return
Authorization number (MRA) first, providing the following infor-
mation:
1. Part Number, Serial Number
2. Name and telephone number of someone who can answer
technical questions related to the product and its application.
3. Return Shipping Address
4. Brief Description of the Symptom
5. Brief Description of the Application
Once you have received a Material Return Authorization number,
ship the product prepaid in its original packing to:
Flowline Factory Service
MRA _____
10500 Humbolt Street
Los Alamitos, CA 90720
To avoid delays in processing your repair, write the MRA on the
shipping label. Please include the information about the malfunc-
tion with your product. This information enables our service tech-
nicians to process your repair order as quickly as possible.
WARRANTY
Flowline warrants to the original purchaser of its products that such
products will be free from defects in material and workmanship under
normal use and service for a period which is equal to the shorter of
one year from the date of purchase of such products or two years from
the date of manufacture of such products.
This warranty covers only those components of the products which
are non-moving and not subject to normal wear. Moreover, products
which are modified or altered, and electrical cables which are cut to
length during installation are not covered by this warranty.
Flowline's obligation under this warranty is solely and exclusively
limited to the repair or replacement, at Flowline's option, of the prod-
ucts (or components thereof) which Flowline's examination proves to
its satisfaction to be defective. FLOWLINE SHALL HAVE NO
OBLIGATION FOR CONSEQUENTIAL DAMAGES TO PERSON-
AL OR REAL PROPERTY, OR FOR INJURY TO ANY PERSON.
This warranty does not apply to products which have been subject to
electrical or chemical damage due to improper use, accident, negli-
gence, abuse or misuse. Abuse shall be assumed when indicated by
electrical damage to relays, reed switches or other components. The
warranty does not apply to products which are damaged during ship-
ment back to Flowline's factory or designated service center or are
returned without the original casing on the products. Moreover, this
warranty becomes immediately null and void if anyone other than ser-
vice personnel authorized by Flowline attempts to repair the defective
products.
Spec Tech Industrial 203 Vest Ave. Valley Park, MO 63088 Phone: 888 SPECTECH
E-mail: sales@spectechind.com
Symprobe™ RF Capacitance
Level Transmitter
Model LP75
Owner's
Manual
Version 4.1A
© 2005 FLOWLINE Inc.
All rights reserved.
Manual # LC900006
05/05
Products which are thought to be defective must be shipped prepaid
and insured to Flowline's factory or a designated service center (the
identity and address of which will be provided upon request) within
30 days of the discovery of the defect. Such defective products must
be accompanied by proof of the date of purchase.
Flowline further reserves the right to unilaterally wave this warranty
and to dispose of any product returned to Flowline where:
a. There is evidence of a potentially hazardous material present
with product.
b. The product has remained unclaimed at Flowline for longer than
30 days after dutifully requesting disposition of the product.
THERE ARE NO WARRANTIES WHICH EXTEND BEYOND
THE DESCRIPTION ON THE FACE OF THIS WARRANTY. This
warranty and the obligations and liabilities of Flowline under it are
exclusive and instead of, and the original purchaser hereby waives, all
other remedies, warranties, guarantees or liabilities, express or
implied. EXCLUDED FROM THIS WARRANTY IS THE IMPLIED
WARRANTY OF FITNESS OF THE PRODUCTS FOR A PARTIC-
ULAR PURPOSE OR USE AND THE IMPLIED WARRANTY OF
MERCHANT ABILITY OF THE PRODUCTS.
This warranty may not be extended, altered or varied except by a writ-
ten instrument signed by a duly-authorized officer of Flowline, Inc.
www.spectechind.com
®

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Symprobe LP75 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for FlowLine Symprobe LP75

  • Page 1 Flowline’s obligation under this warranty is solely and exclusively 30 days after dutifully requesting disposition of the product. limited to the repair or replacement, at Flowline’s option, of the prod- ucts (or components thereof) which Flowline’s examination proves to THERE ARE NO WARRANTIES WHICH EXTEND BEYOND its satisfaction to be defective.
  • Page 2: Specifications

    Classification: General purpose CE Compliance: EN 61326 EMC 3/4" NPT (3/4" G) Standard Probe Lengths 3.5" (89mm) 4.8" (122mm) Maximum Temperature/Voltage Flowline Electrical Loading Limits Derating @ Continuous 20 mA 1,400 1,200 Unacceptable Unacceptable 1,000 Range Range Acceptable Acceptable Range...
  • Page 3: Safety Precautions

    User’s Responsibility for Safety: and a guard wire which are completely encapsulated in FEP. The FLOWLINE manufactures a wide range of liquid level sensors and Symprobe generates a 1 MHz pulse-wave radio frequency signal from technologies. While each of these sensors are designed to operate the continuous capacitance probe.
  • Page 4: Installation

    Side wall installations require a side mount bracket from 1. Wiring to a FLOWLINE Continuous Controller (LC52): FLOWLINE, model LM50-1001, and a 2" to 3/4" NPT reducer bush- Connect the (+) terminal to the positive 24 VDC, 4-20 mA termi- ing or equivalent.
  • Page 5: Standard Calibration

    CALIBRATION CALIBRATION Step Six Step Seven The Symprobe™ is factory calibrated over the entire range of the 1. Power LED: This Green LED lights when DC power is On. 2. Input terminal: Connect the output device to this terminal. Note probe.
  • Page 6: Maintenance

    - 60 - 70 % of Span FLOWLINE recommends re-calibrating the LP75 transmitter when: 1) the application medium has a low conductivity or dielectric value; 2) the product will be applied in several different mediums with widely differing conductivity and/or dielectric values.