Configuration Of Analog Sensors; Type 0 - 1/10/30 Volt And 0/4 - 20 Ma - Beko METPOINT BDL Instructions For Installation And Operation Manual

Graphic display-recorder
Table of Contents

Advertisement

14.2.2.8 Configuration of analog sensors

Short overview of the possible
Except for BEKO-Digital, in this case see Chapter
Digital sensor type)
and
The
alarm
(see Chapter
measuring
data), the
resolution
Chapter
12.2.2.2 Denoting the measuring data and determining the resolution of the decimal
described in Chapter
12.2.2 Sensor
For the marking of the text fields, see Chapter
14.2.2.8.1 Type 0 - 1/10/30 volt and 0/4 – 20 mA
Main menu  Settings  Sensor settings  C3  Type text field  0 - 1/10/30 V
METPOINT® BDL
Type
settings including examples.
12.2.2.6 Dew point sensor with the BEKO-DIGITAL
12.2.2.4 Alarm
settings)
of the decimal places and the
settings.
12.2.2.7 Marking and setting text
12.2.2.1 Selection of the sensor type (Example BEKO-
and
recording
buttons (see Chapter
short name
Scaling of the sensor (here, for example,
type 0 – 10 V - corresponds to 0 – 250°C)
can be found in the data sheet of the con-
nected sensor.
For
scale 0 V,
please enter the lower scale
value and for
scale 10 V
value.
The
ext. Sensor supply voltage
on when the sensor requires this.
With the
Set-value-to
measured data of the sensor can be set to
a certain value.
The positive or negative difference of the
offset
is indicated.
With the
reset
button, the
reset to zero.
Operation BDL
type.
12.2.2.3 Recording
or the
value name
(see
places) are all
fields!
the upper scale
is switched
button (offset), the
offset
can be
49