Operation BDL
Main menu Settings Sensor settings A1
Main menu Settings Sensor settings A1
See also Chapter
12.2.2.7 Marking and setting text fields
Note:
After having confirmed by
accepted.
Caution:
Reference temperature and reference pressure (setting ex works 20°C, 1000 hPa):
All the volume flow (m³/h) and consumption values (m³) that are indicated on the display refer to 20°C
and 1000 hPa (according to ISO 1217 suction condition).
Alternatively, 0°C and 1013 hPa (=standard cubic metre according to DIN 1343) can also be entered
as the reference. Under no circumstances must the operating pressure or the operating temperature
be entered into the reference conditions!
40
OK,
the lettering is switched to black again. The values and settings have been
Now, a
name
can be entered and, at an ex-
change of the sensor, the
the previous sensor.
Subsequent to marking and confirming by OK,
the configuration of the sensor is completed.
counter reading
of
METPOINT® BDL