GE RSDetection RS-S131-200-ER0000 User Manual page 87

Table of Contents

Advertisement

g
Parameter
Support
ZLD
No
ZMD
No
ZHD
No
Electrometer Gain
RLN
No
RMN
No
RHN
No
RLV
No
RMV
No
RHV
No
General
UID
Yes
HEU
Yes
DTF
Yes
WDE
No
COM1
C1B
Yes
C1P
Yes
C1H
Yes
COM2
C2B
Yes
C2P
Yes
Copyright © 2016 General Electric Company. All Rights Reserved
Returns 0.
Returns 0.
Returns 0.
Returns 0.
Returns 0.
Returns 0.
Returns 0.
Returns 0.
Returns 0.
Sets or returns the unit address 0-Z except '@'.
Sets or returns the R to Sv conversion factor..
Legacy date format 0 = mm/dd/yy, 1=dd/mm/yy.
Watchdog enable. Returns 1.
Sets or returns the RS=232 port baud rate. Valid values are:
300,600,1200,2400,9600,19200. User must restart the unit after
this command is sent for the changes to go into effect. Make
sure COM1 is available on the unit before sending this
command.
Parity for the RS-232 port. Valid values are: (E, O, N). User must
restart the unit after this command is sent for the changes to
go into effect. Make sure COM1 is available on the unit before
sending this command.
Handshaking for the RS-232 port. Valid values are:
0 = No handshake, 1=RTS/CTS, > 1 delay in mSec. User must
restart the unit after this command is sent for the changes to
go into effect. Make sure COM1 is available on the unit before
sending this command.
Sets or returns the RS=232 port baud rate. Valid values are:
300,600,1200,2400,9600,19200. User must restart the unit after
this command is sent for the changes to go into effect.
Parity for the RS-232 port. Valid values are: (E, O, N). User must
RSDetection
User's Manual
Comments
S131-200-UM Rev A
Page 73

Hide quick links:

Advertisement

Table of Contents