B : Operation Instructions:
B-1 : Power on the instrument by pressing POWER KEY,
and then press 1Φ3Φ KEY to select the 1Φ3W system,At the
time, display will show 1Φ3W connect diagram approx 4 sec.,
and the selected name of system will be appeared on bottom
left display of screen 2.
B-2 : Connect the line voltage L1, L2 and Vn (Neutral) to
V1, V2 and N terminals of the instrument.
B-3 : Place the conductor of CP-1201(A1), CP-1201(A2)
hook to A1 and A2 as screen 1.
B-4 : Connect the outputs of clamp meter
CP-1201(A1)、CP-1201(A2) to A1 and A2 terminals of
the instrument.
B-5 : The related measuring factors will be appeared on
display, about the instruction of factor please refer
appendix 1 ( 5-16, page 68 ).
Remark:
During connection ,If want to check connect Correct or not,
please press and hold the "▼" key, display will show 1Φ3W
connect diagram,when release the "▼" key , will back to
measurement.