Active (Real) Power
DW-6195 + CP1201
Range
0.000 to 9.999 KW
10.00 to 99.99 KW
100.0 to 999.9 KW
1.000 to 9.999 MW
* The resolution is changed according the different ACA range.
Apparent Power
DW-6195 + CP1201
Range
0.000 to 9.999 KVA
10.00 to 99.99 KVA
100.0 to 999.9 KVA
1.000 to 9.999 MVA
* The resolution is changed according the different ACA range.
Reactive Power
DW-6195 + CP1201
Range
0.000 to 9.999 KVAR
10.00 to 99.99 KVAR
100.0 to 999.9 KVAR
1.000 to 9.999 MVAR
* The resolution is changed according the different ACA range.
Remark :
*
When the Reactive power value ( Q1 to Q3 ) show " - "
indicator, it means the " current phase " lag than the " voltage
phase ", the load character is induction.
*
When the Reactive power value ( Q1 to Q3 ) do not show " - "
indicator, it means the " current phase " lead than the " voltage
phase ", the load character is capacitance.
Resolution
*0.001/0.01/0.1 KW
*0.01/0.1 KW
0.1 KW
0.001 MW
Resolution
*0.001/0.01/0.1KVA
*0.01/0.1 KVA
0.1 KVA
0.001 MVA
Resolution
*0.001/0.01/0.1KVAR
*0.01/0.1 KVAR
0.1 KVAR
0.001 MVAR
7
System
Accuracy
±(1.2%+0.008KW)
±(1.2%+0.08KW)
±(1.2%+0.8KW)
±(1.2%+0.008MW)
System
Accuracy
±(1.2%+0.008KVA)
±(1.2%+0.08KVA)
±(1.2%+0.8KVA)
±(1.2%+0.008MVA)
System
Accuracy
±(1.2%+0.008 KVAR)
±(1.2%+0.08 KVAR)
±(1.2%+0.8 KVAR)
±(1.2%+0.008 MVAR)