3 phase power analyzer
Table of Contents

Advertisement

ACA
Range Resolution
20A
0.001A,
0.01A,
200A
0.01A,
0.1A,
1200A 0.1A,
1A,
Remark :
*
When the Active power value ( P1 to P3 ) and Apparent
power value ( S1 to S3 ) show " - " indicator, it means the
current probe is under the reverse direction that compare
with the real measuring current.
Power factor
Range
Resolution
0.00 to 1.00
0.01
Remark :
* PFH : Long term power factor
* PFΣ:
For 3Φ4W, 3Φ3W, 1Φ3W : PFΣ = PΣ/SΣ
For 1Φ2W : PF1 = P1/S1
Φ ( Phase angle )
Range
Resolution
-180° to 180°
0.1°
Frequency
Range
Resolution
45 to 65 Hz
0.1 Hz
Accuracy
< 10 A
Meter + CP-1201
10 A
Meter only
< 100 A
Meter + CP-1201
100 A
Meter only
< 1000 A
Meter + CP-1201
1000 A
Meter only
6
±(1 %+0.1A)
±(0.5 %+0.02A)
±(1 %+0.5A)
±(0.5 %+0.2A)
±(1 %+5A)
±(0.5 %+2A)
Accuracy
±0.04
Accuracy
±1°
* COS ( PF )
Accuracy
±0.1 Hz