Periodic Inspection; Preventative Maintenance Testing - Steris gi4000 User And Maintenance Manual

Electrosurgical unit
Table of Contents

Advertisement

13.1 Periodic Inspection

The gi4000 should be visually inspected at least every six (6) months. This inspection should
include checks for:
1.
Damage to the gi4000.
2.
Damage to the power cord.
3.
Damage to the power cord receptacle or the cord lock.
4.
Tightness of the power cord plug.
5.
Loose or missing hardware or screws.
6.
Damage to the sterile water bottle holder.
7.
Damage to the argon gas bottle mounting or difficulty screwing in the argon gas
bottle.
8.
Cleanliness and proper fit and function of the patient receptacles.
9.
Accumulation of lint or debris on the vent ports or within the unit.
10.
Proper functionality of the front panel indicator lights: all illuminate and cycle upon
startup.
11.
Proper functionality of the standby button on the footswitch and touchscreen.

13.2 Preventative Maintenance Testing

NOTE: The gi4000 ESU should undergo preventive maintenance testing annually.
NOTE: When performing electrical testing which requires the use of test leads, test
leads should be fully stretched out and not touching each other or any conductive
surfaces, including the floor.
A thorough preventative maintenance program is essential to safe and proper equipment
operation. Comprehensive instructions for annual preventative maintenance can be found in
this section of the manual. STERIS Endoscopy recommends this preventative maintenance
test be performed on an annual basis.
Each unit is supplied with a serialized Product Test Data Sheet, which tabulates the results of
the factory tests that were performed on the unit. This data may be used as a reference for
subsequent tests and should be made available to the biomedical technician conducting the
preventative maintenance testing.
Visit www.steris.com to download 00732974, Preventative Maintenance Test Data
Summary Sheet for the gi4000 ESU which is required to perform the gi4000 annual
maintenance inspection.
00732467 Rev. D
Page 84 of 110