Page 4
Complete instructions for uncrating have been furnished. If missing, contact STERIS for a replacement copy, giving the serial, equipment and model numbers of the unit. STERIS carries a complete line of accessories for use with this table. STERIS representatives will gladly review these with you. Advisory...
Page 5
STERIS maintains a global staff of well equipped, factory-trained technicians to provide these services, as well as expert repair services.
Page 6
STERIS Limited STERIS House Jays Close Viables Basingstoke Hampshire RG22 4AX UNITED KINGDOM Manufactured by: STERIS Corporation 2720 Gunter Park East Montgomery, AL 36109 • USA TEL: 334 277 6660 FAX: 334 271 5450 www.steris.com Class 1 Equipment Type B Equipment...
TABLE OF CONTENTS Section Number Description Page Safety Precautions ........................1-1 Symbols on Table ..........................1-5 Symbols on Hand Control ........................1-6 Important User Information ......................2-1 Pinch Point Warnings ..........................2-1 Patient Positioning and Weight Limitation ..................... 2-3 Patient Safety Straps ..........................2-4 General Description ..........................
Page 8
TABLE OF CONTENTS (CONT’D) Section Number Description Page Cleaning Table............................5-2 5.2.1 General ............................5-2 5.2.2 After Each Usage ......................... 5-3 5.2.3 End-of-Day Cleaning Procedure ....................5-4 5.2.4 Weekly Cleaning Procedure......................5-4 Bi-Weekly Maintenance ......................... 5-4 Monthly Maintenance..........................5-4 Battery Charging Procedure ........................5-5 Preventive Maintenance Schedule ......................
Page 9
Figure 2-6. STERIS 4085 General Surgical Table Image Amplification Coverage ..........2-9 Figure 3-1. Install/Remove Battery Fuse ......................3-2 Figure 3-2. Connect STERIS 4085 General Surgical Table to Facility Power (Typical) ........3-3 Figure 3-3. Connect Hand Control to STERIS 4085 General Surgical Table (Typical)......... 3-4 Figure 3-4.
SAFETY PRECAUTIONS The following Safety Precautions must be observed when operating or servicing this STERIS 4085 General Surgical Table. WARNING indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. For emphasis, certain Safety Precautions are repeated throughout the manual. It is important to review ALL Safety Precautions before operating or servicing the unit.
Page 12
WARNING – INSTABILITY HAZARD: Possible patient or user injury, as well as table or accessory failure, may result from using STERIS table accessories for other than their stated purpose – or from using accessories manufactured and sold by other companies on STERIS surgical tables.
Page 13
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD: Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service personnel. Maintenance performed by unqualified personnel or installation of unauthorized parts could cause personal injury, result in improper equipment performance, invalidate the warranty, or result in costly damage.
Page 14
Route the hand control cord clear of any pinch points where cord could be damaged. Use of incorrect hydraulic oil may severely damage the table and/or cause malfunction. Contact STERIS for proper hydraulic oil. After performing cleaning procedures, ensure pads, tabletop and X-ray tops are completely dry before reinstalling.
1.1 Symbols on Table The following is a key to symbols on the STERIS 4085 General Surgical Table. Table 1-1. Definition of Symbols on Surgical Table Symbol Definition Type B Equipment Protective Earth Ground Equipotentiality Electric Shock Hazard Control Power ON...
Powered By Battery 1.2 Symbols on Hand Table 1-2 and Table 1-3 are keys to the symbols appearing on the STERIS 4085 General Surgical Table Hand Control and Hand Control Control LCD Display. Table 1-2. Definition of Symbols on Hand Control...
Table 1-2. Definition of Symbols on Hand Control (Cont’d) Symbol Definition HEIGHT DN Tabletop Down (Height Up; Lower) TREND Move Tabletop into Trendelenburg REVERSE TREND Move Tabletop into Reverse Trendelenburg FLOOR Floor Lock Function (see Table 1-3) LOCK Slide Tabletop Toward Head End (SLIDE HEAD) Slide Tabletop Toward Foot End (SLIDE FOOT) Table 1-3.
Page 18
Table 1-3. Definition of Hand Control LCD Display Screen Icons (Cont’d) Symbol Definition Floor Locks Unlocked (Orange). ALS Notification (ICON and Text). Maintenance Notification (Blue). Kidney Bridge (Blue). Keypress (Green). Slide in Head Direction (Blue). Slide in Foot Direction (Blue). Slide Displayed in Inches (Blue).
When inertia-driven or power-driven table parts close (especially during extreme tabletop articulation), pinch-point hazards exist (see Warnings Figure 2-1). All personnel involved in tabletop positioning should examine and be aware of these points before operating the STERIS WARNING – PINCHING 4085 General Surgical Table. HAZARD: Pinch points shown (circled) in Figure 2-1: •...
Do not exceed the lowest weight limit, table or and Weight Limitation accessory. WARNING – TIPPING The STERIS 4085 General Surgical Table is designed to safely HAZARD: support (with patient in normal orientation): 1100 lb (499 kg) without patient posturing or tabletop slide (except raise/lower); 1000 lb •...
2.4 General The STERIS 4085 General Surgical Table (see Figure 2-3): Description • is a mobile, electrohydraulically operated surgical table specifically designed to provide the complete patient positioning CAUTION –...
Table. and found in compliance with IEC 60601-1-2:2001, Medical • The STERIS 4085 General Surgical Table should not be used Electrical Equipment – Part 1; adjacent to or stacked with other equipment. If so, table should be General Requirements for observed to verify normal operation.
7. Seat Section (Sliding) 17. Service Bar Code Label 8. Leg Section 18. Primary Fuses 9. Power Panel 19. Main Power Switch 10. Power Cord 20. AC Receptacle Figure 2-3. STERIS 4085 General Surgical Table Components (Typical) 150832-639 Operator Manual Important User Information...
100-240 Vac, One-Phase, 50/60 Hz, 4.0 Amp NOTE: Each table is shipped from the factory configured to the electrical requirement specified on the factory order. If this electrical configuration needs to be changed in the field,consult STERIS for the needed procedure and required materials. 2.5.6 Environmental Temperature: 32-122°F (0-50°C)
INSTALLATION INSTRUCTIONS Check the overall condition of the STERIS 4085 General Surgical WARNING – PERSONAL Table and any applicable accessories upon receipt to ensure INJURY HAZARD: nothing was damaged during shipment. If damage appears to have occurred, contact the shipping company and STERIS.
Figure 3-1. Install/Remove Battery Fuse NOTE: A green Main Power Switch indicates facility power is available to the STERIS 4085 General Surgical Table. A lighted plug symbol on the LED Display indicates the batteries are charging. Refer to S ECTION 5.5, B...
Ensure caps cover SERVE Foot End Head End Figure 3-2. Connect STERIS 4085 General Surgical Table to Facility Power (Typical) Install head section as follows: a. Fully loosen both locking thumbscrews located at bottom edge of back frame (see Figure 3-6).
Cover Control Panel Click Figure 3-3. Connect Hand Control to STERIS 4085 General Surgical Table (Typical) b. Insert both leg section installation rods into seat frame until fully engaged. Listen for "click" sound to indicate correct installation (see Figure 3-8).
Figure 3-4. Connect Panel Protective Cover Control Ensure caps cover sockets when not in use. Panel SERVE Click Disconnect Foot Control Connect Foot Control Figure 3-5. Connect Optional Foot Control to STERIS 4085 General Surgical Table (Typical) Installation Instaructions Operator Manual 150832-639...
Figure 3-6. Install Head Section LCD Display Tabletop Positioning Primary Hand Control Symbols Figure 3-7. Hand Control LCD Display WARNING – PERSONAL «Click» INJURY HAZARD: The table leg section is designed to support the patient legs. Do not put upper body weight on the leg section.
OPERATING INSTRUCTIONS 4.1 Operating Table Operate the STERIS 4085 General Surgical Table as follows: 1. Review Pinch Point Warnings as presented in S 2.1, P ECTION INCH WARNING – PERSONAL of this manual. OINT ARNINGS INJURY HAZARD: ® STERIS 4085 General Surgical Table is equipped with INTELLIPOWER •...
NOTE: A patient grounding post/potential equalization terminal (male chest compressions. connector, DIN 42801) is provided. The mating female connector for WARNING – PERSONAL patient grounding is not furnished by STERIS. INJURY HAZARD: The Table Leg Section is designed to support the patient legs. Do not put upper body weight on the Leg Section.
The hand control is reactivated by depressing ANY button but STOP. The STERIS 4085 General Surgical Table is equipped with a hand control (see Figure 4-3). This hand control contains easy-to-read ICONs and illuminates the key pad for easy articulation identification when room is dark.
LCD Display Figure 4-3. STERIS 4085 General Surgical Table Primary Hand Control SLIDE FOOT: When depressed, Zip-Slide moves tabletop toward foot end. When slide returns to central ("0") position, movement stops for one second before continuing. LCD Display Slide Head ICON (refer to Table 1-3) displays.
Page 38
BACK UP: When depressed, the tabletop back section raises. BACK DN: When depressed, the tabletop back section lowers. LEG UP: When depressed, the tabletop leg section raises. LEG DN: When depressed, the tabletop leg section lowers. KIDNEY UP: When depressed, the tabletop Kidney™ Elevator Mechanism raises to the maximum 4"...
4.2.2 Hand Control LCD The Hand Control is equipped with a LCD (Liquid Crystal Display) Display Description Display to assist in the table operation (see Figure 4-4). The LCD Display indicator descriptions are as follows: NOTE: Numbers in ( ) refer to items shown in Figure 4-4. LCD Display (1): displays any control system messages to the user about table operation.
The hand control is reactivated by depressing ANY button but STOP. 4.4 Optional Foot The STERIS 4085 General Surgical Table may be operated using the optional foot control as follows: Control Operation 1. Connect foot control cord to appropriate control panel socket on table column (refer to Figure 3-5).
Figure 4-5. Backup (Override) Control System 4.5.2 Backup Hand Control The Backup Hand Control pushbutton descriptions are as follows (Figure 4-6): Description NOTE: Simply release the Backup Hand Control pushbutton when desired STERIS 4085 General Surgical Table Tabletop position has 150832-639 Operator Manual Operating Instructions...
Page 42
been reached and table automatically stops and locks in position. Refer to S 2.5.2, R , for the positioning of the ECTION ANGE OF OTION following table functions. SLIDE HEAD: When depressed, Zip-Slide moves tabletop toward head end. When slide returns to central ("0") position, movement stops for one second before continuing.
FLOOR LOCK UNLOCK Figure 4-6. Backup (Override) Hand Control 4.6 Leg Section The STERIS 4085 General Surgical Table leg section is designed to ™ Removal be easily removed (Hi-Lock locking mechanism) to enable procedures requiring optional leg supports. Remove the leg section as follows: 1.
Handle Figure 4-7. Leg Section Removal 4.7 Table Pads The conductive Tabletop Pads (see Figure 4-8) are backed with fastening strips which fasten to mated strips on the tabletop. To install, proceed as follows: 1. Place pads in position and firmly press fastening strips together. 2.
4.8 X-Ray Top The STERIS 4085 General Surgical Table is designed to enable installation of an X-ray top to allow use of X-ray cassette film. Attach Installation the four-part X-ray top as follows (see Figure 4-9): 1. Carefully remove X-ray top sections from packing container.
4.9 Accessories/Side The standard STERIS permanently attached 3/8" wide x 1-1/8" tall side rails allow for the use of many standard surgery table Rails attachments and accessories. The rails consist of one rail mounted to each side of each tabletop main section.
Contact STERIS for details. performed by unqualified Maintain a record of all maintenance procedures performed on this personnel or installation of STERIS 4085 General Surgical Table. If a problem occurs, refer to unauthorized parts could 6, T , or contact STERIS.
Disinfecting/Deoderizing/ INJURY HAZARD: When Cleaning Wipes cleaning/disinfecting table, do not use phenolics, which may cause patient skin burns if * Contact STERIS for ordering information. inadequately rinsed off, or alcohol, which does not have sufficient cleaning/ disinfection properties. 5.2.1 General...
5.2.2 After Each Usage After each use of the STERIS 4085 General Surgical Table, clean/ disinfect as follows: WARNING – INFECTION NOTE: Ensure all protective covers are installed over any open HAZARD: To protect against receptacles. aerosols being reflected from potentially contaminated 1.
S 5.2.2, A Procedure ECTION FTER SAGE 5.2.4 Weekly Cleaning After each weekly use of the STERIS 4085 General Surgical Table, clean/disinfect table as follows: Procedure 1. Perform Steps 1 through 4 of S 5.2.2, A ECTION FTER SAGE 2.
Power Panel. Lead acid batteries last longer if not fully discharged. Therefore, to obtain the longest life and capacity from your STERIS 4085 General Surgical Table batteries, always connect ac power cord to table base and plug into an appropriate ac receptacle as often as possible, and as long as possible, with the Main Power Switch set to ON (lighted green).
table operation without damaging the batteries. However, when the last green diode bar is OFF, batteries must be recharged without further delay or battery damage occurs. If all diodes are OFF, batteries are completely drained and the table can only operate on main facility power.
ROUBLESHOOTING equipment damage. NOTE: Never permit unqualified persons to service the table and Preventive Maintenance is not covered under warranty. Table 5-2. Preventive Maintenance Schedule for STERIS 4085 General Surgical Table Service Required Minimum Frequency 1.0 PREPARATION FOR PREVENTIVE MAINTENANCE Discuss equipment operation with department personnel.
Table 5-2. Preventive Maintenance Schedule for STERIS 4085 General Surgical Table (Cont’d) Service Required Minimum Frequency Verify condition of batteries and battery charger status. 1 x per year Verify Auto Limit Sensor status. 1 x per year 6.0 TABLETOP OPERATION Check operation of Level function.
Pad, Back/Seat Section 542200240 Pad, Leg Section 542200241 5.8 ALS Check The STERIS 4085 General Surgical Table is equipped with the Auto Limit Sensor system. This system provides instant user feedback on ™ any conflicting tabletop articulations. An Auto Limit Sensors (ALS) WARNING –...
KIDNEY Elevator and lateral tilt. b. Trendelenburg and back. c. Leg. 4. Use a level and ensure tabletop is level. If not, call STERIS. 5.10 Auxiliary Control Check the STERIS 4085 General Surgical Table auxiliary control system as follows: Check 1.
ECTION OUTINE AINTENANCE • Failure to perform periodic STERIS. Service charges may be incurred, consult your warranty for inspections of the table details. A trained service technician will promptly place your STERIS could result in serious 4085 General Surgical Table in proper working order.
STOP) is pressed . Hand control not plugged in properly – correct. Faulty hand control** – replace. Call STERIS.* Table remains on battery power even when main Incorrect power cord connection – correct. power cord is plugged into facility power.
Table 6-2. Troubleshooting Chart (LCD Message) FAULT DISPLAYED FAULT DESCRIPTION FAULT ACTION 1 KEY ONLY Two pushbuttons are pressed at the same Press one button only time DUAL COMMAND Commands are sent from the optional foot Hand control has priority pedal and the hand control at the same time.
(Cu), and copper alloys (Cu/x), plastic, synthetic rubber, and/or cause malfunction. plating (Cr, Ni, Zn, Au), and adhesives not known to require special Contact STERIS for proper disposal methods at date of this manual. hydraulic oil. Hydraulic Oil: contained in hydraulic components located in the base, column, leg section, back section, and hydraulic system lines and hoses.
Page 62
, for a list of terminology used in this document. EFINITIONS Thank you for choosing this fine STERIS product. INDICATIONS FOR USE STERIS is committed to ensuring your continued The STERIS 4085 General Surgical Table is designed for satisfaction. This manual contains...
Page 63
EC Authorized Representative STERIS Limited STERIS House Jays Close Viables Basingstoke Hampshire RG22 4AX UNITED KINGDOM Manufactured by: STERIS Corporation 2720 Gunter Park East Montgomery, AL 36109 • USA Class 1 Equipment Type B Equipment Classified as IPX4 (Splash-proof) Equipment not suitable for use...
Page 64
TABLE OF CONTENTS Section Number Description Page Safety Precautions ........................1-1 Terms, Definitions, Symbols and Special Tools................. 2-1 Terms and Definitions ..........................2-1 Symbols ..............................2-1 Special Tools/Materials Required for Maintenance ................2-2 Principles of Operation ......................... 3-1 General ..............................3-1 Table Overview .............................
Page 65
TABLE OF CONTENTS (CONT’D) Section Number Description Page 4.10 Leg Section Checks ..........................4-8 4.11 Replacement Parts and Supplies ......................4-8 Troubleshooting ..........................5-1 General ..............................5-1 Electrical Troubleshooting ........................5-7 Field Test Procedure ........................6-1 General ..............................6-1 Test Instrumentation Needed ........................ 6-1 Table Level Calibration ..........................
Page 66
TABLE OF CONTENTS (CONT’D) Section Number Description Page Floor Lock Cylinder Replacement......................7-6 Floor Lock Foot (Leveling Mount) Replacement ..................7-6 7.10 Caster Replacement ..........................7-7 7.11 Floor Lock Manifold Assembly Replacement..................7-7 7.12 Hydraulic Pump Replacement ....................... 7-7 7.13 Auxiliary (Backup) Hand Control Replacement..................7-8 7.14 Table Column Cover Cap Replacement ....................
Page 69
Table 2-4 Tools for STERIS 4085 General Surgical Table Maintenance ............2-6 Table 3-1 STERIS 4085 General Master Control and Hand Control Assembly Compatibility Matrix....3-7 Table 4-1 Preventive Maintenance Schedule for the STERIS 4085 General Surgical Table......4-2 Table 4-2 STERIS 4085 General Surgical Table Replacement Parts .............
Section 1: Safety Precautions The following Safety Precautions must be observed when operating or servicing the STERIS 4085 General Surgical Table. WARNING indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. For emphasis, certain Safety Precautions are repeated throughout the manual. It is important to review ALL Safety Precautions before operating or servicing the unit.
Page 71
Do not store items on table base. Doing so may result in equipment damage or inadvertent tabletop movement placing the patient and/or user at risk of injury. Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service personnel. Maintenance performed by unqualified personnel or installation of unauthorized parts could cause personal injury, result in improper equipment performance, invalidate the warranty or result in costly damage.
Page 72
Route the hand control cord clear of any pinch points where cord could be damaged. Use of incorrect hydraulic oil may severely damage the table and/or cause table malfunction. Contact STERIS for proper hydraulic oil. After performing cleaning procedures, ensure pads and tabletop are completely dry before reinstalling.
Page 73
CAUTION – POSSIBLE EQUIPMENT DAMAGE (CONT’D): When the bearing assembly is removed from the table, the seat section side member becomes free- floating. Be careful not to damage the slide potentiometer pulley that sits freely in the groove in the seat section side member.
Surgical Table ™ Surgical Table Tabletop. Pads. Surgical Table - STERIS 4085 General Surgical Table. Flex - Tabletop position where back section moves down Table Base - Painted welded steel covered by stainless- and Tabletop moves into Reverse Trendelenburg. steel or plastic enclosure. Base contains casters, floor...
Table 2-4 shows tools and packs necessary to repair very important. and maintain the STERIS 4085 General Surgical Table. Table 2-1. Definition of Symbols on Surgical Table Symbol Definition...
Page 76
Table 2-1. Definition of Symbols on Surgical Table (Cont’d) Symbol Definition Frequency Rating of Unit Power Rating Serial Number of the Unit IPX4 Splash-Proof Equipment INTELLIPOWER ® Dual Power System Power Panel (Battery Charge/Discharge Status) CHARGE BATTERY INTELLI POWER Powered By Battery Table 2-2.
Page 77
Table 2-2. Definition of Symbols on Hand Control (Cont’d) Symbol Definition TREND Move Tabletop into Trendelenburg REVERSE TREND Move Tabletop into Reverse Trendelenburg FLOOR Floor Lock Function (see Table 2-3) LOCK Slide Tabletop Toward Head End (SLIDE HEAD) Slide Tabletop Toward Foot End (SLIDE FOOT) Table 2-3.
Page 78
Table 2-3. Definition of Hand Control LCD Display Screen Icons (Cont’d) Symbol Definition Maintenance Notification (Blue). Kidney Bridge (Blue). Keypress (Green). Slide in Head Direction (Blue). Slide in Foot Direction (Blue). Slide Displayed in Inches (Blue). Slide Displayed in Centimeters (Blue). 764333-766...
Page 79
Table 2-4. Tools for STERIS 4085 General Surgical Table Maintenance Part Number Tool Quantity P752911-091 Wrench - Socket Set, Inch and Metric, Deep Well P764324-655 Wrench - Allen, Metric, Ball P764331-416 Wrench - Ratcheting, Metric P764329-992 Pack, Velcro, 5 Feet (With Instructions)
The following sections give a detailed description of the STERIS 4085 General Surgical Table. P RINCIPLES OF The STERIS 4085 General Surgical Table is powered by covers power distribution, tabletop PERATION either internal battery or facility electric power through articulation and floor lock operation in detail.
4) The ALS System is NOT active during Backup The STERIS 4085 General Surgical Table may be operated Control operation. using the optional Foot Control as follows: 5) During Backup Control operation, operation with 1.
3.3.2.7 Floor Lock Operation are depleted. The master computer selects which outputs are to be NOTE: Not all STERIS 4085 General Surgical Tables are actuated based on inputs from the hand control, foot equipped with compensating floor locks. 764333-766...
Processor) is the heart of the master controller, with (P150832-769) may be configured for either the software stored internally using Flash ROM. Solenoid Cmax 110 NA, the 4085 General, the 5085 General outputs are generated by open-collector drivers. A or the 5085 SRT Surgical Table by DIP switch watchdog timer and appropriate software monitor I/O settings.
Page 84
d. Other fuses internal to the power supply are as board are mounted on a common connector plate follows: on the saddle cover. These provide primary table control functioning with the main hand control or • Fuse F1 protects the battery charging circuit surgeon pedal and the e-Serve port provides the for Battery 2+ at TB2.
control pass through the table control board at 2. When facility power is not connected (cord not P5 but bypass the microprocessor control. Thus, connected or power switch is open): even with minimal primary circuit functions • Line Power LED below the bar graph is OFF. available, the backup system is likely to be •...
When the system is operating on bat- tery power, the Hand Control icon indicates the battery charge level in three stages: Full, Half and Empty. Table 3-1. STERIS 4085 General Master Control and Hand Control Assembly Compatibility Matrix Hand Control* Master Control...
In addition to the routine maintenance, regularly 4.3 CLEANING TABLE scheduled preventive maintenance is essential for safe and reliable operation of the STERIS 4085 General 4.3.1 General Surgical Table. Procedures described in this section of the Maintenance Manual should be performed at WARNING –...
Page 88
Table 4-1. Preventive Maintenance Schedule for the STERIS 4085 General Surgical Table Minimum Service Required Frequency 1.0 PREPARATION FOR PREVENTIVE MAINTENANCE Discuss equipment operation with department personnel. 1x per year Remove pads. Examine pad covers and fastening strips (pads and table).
Page 89
Table 4-1. Preventive Maintenance Schedule for the STERIS 4085 General Surgical Table Minimum Service Required Frequency Check all cables for damage or fraying. 1x per year Inspect electrical cord and Hand Control cable. 1x per year Verify operation of Backup (Override) Hand Control.
5. Unplug hand control. Note hand control plug is locked, do not simply pull on plug. After each use of the STERIS 4085 General Surgical 6. Clean hand control as follows: Table, clean/disinfect as follows: NOTE: Do NOT immerse hand control in liquid.
Main Power Switch is set to ON (lighted green) and the ac power is connected to the STERIS • When table is on battery power and primary hand 4085 General Surgical Table. The PLUG icon appears...
5. Insert fuse in cap. Therefore, to obtain the longest life and capacity from your STERIS 4085 General Surgical Table batteries, 6. Insert fuse (with cap) into fuse holder assembly in always connect ac power cord to table base and plug table base.
Figure 4-2. Install/Remove Battery Fuse (Typical) design leg section accessory mounts. 4.7 LEVEL CHECK 4.6 ALS CHECK Check the STERIS 4085 General Surgical Table tabletop WARNING – PERSONAL INJURY AND/OR LEVEL function as follows: EQUIPMENT DAMAGE HAZARD: The Auto 1. Press LEVEL touch pad on Primary Hand Control.
2. Include model number and description as listed in 4.9.2 Floor Lock Cylinder Inspection Table 4-2 or Table 4-3. At any inspection or installation of a STERIS 4085 3. Send your order directly to STERIS Sales and General Surgical Table, inspect the floor lock cylinders Service Center serving your area.
Page 95
Table 4-2. STERIS 4085 General Surgical Table Replacement Parts DESCRIPTION PART NUMBER RECOMMENDED SPARES Primary Hand Control Assembly With 150832-632 Cord Power Cord 093909-354 Fuse (Main), 4.0 Amp 150832-335 Fuse (Set) 342213555 Pad, Drive Head Section 150832-695 Pad, Back/Seat Section...
5.1 GENERAL WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD: This section describes the types of STERIS 4085 General Surgical Table malfunctions or situations that may occur • Repairs and adjustments to this and indicates possible causes and suggested actions for equipment must be made only by remedy.
Page 97
Table 5-1. Troubleshooting Guide Situation Cause Service Instructions 1. No green light on Main Power • Main Power Switch is • Turn Main Power Switch to ON. Switch (if equipped with backlight). OFF. • Plug power cord into appropriate facility •...
Page 98
Table 5-1. Troubleshooting Guide (Cont’d) Situation Cause Service Instructions 6. Table does not function on facility • Control power fuse F2 • Replace control power fuse F2. power (power cord plugged into open. appropriate facility power and battery fuse F8 removed, or battery charge too low) but table does function on battery power.
Page 99
Table 5-1. Troubleshooting Guide (Cont’d) Situation Cause Service Instructions 12. Hand Control displays SENSOR • Bad or loss of sensor • Recalibrate table. FAIL. Level button may or may not calibration data. • Verify battery voltage level and recharge be responsive. as needed.
Page 100
Table 5-1. Troubleshooting Guide (Cont’d) Situation Cause Service Instructions 19. Table has longitudial mechanical • Check lash in rack and • Replace slide motor assembly. play. pinion gear system. • Adjust table for minimum lash. • Check tension in pinion drive chain.
Page 101
Table 5-1. Troubleshooting Guide (Cont’d) Situation Cause Service Instructions 25. Only tabletop slide function • Faulty hydraulic pump. • Verify backup hand control/foot pump operates. system function. • Faulty power supply at pump drive circuit. • Ensure voltage present at pump. •...
Table 5-2 . Troubleshooting (LCD Message) Chart (Cont’d) FAULT DISPLAYED FAULT DESCRIPTION FAULT ACTION BATT FUSE No fuse or blown fuse. Check fuse; plug power cord into facility power. BATT DEF Battery voltages differ by more than Check voltage of each battery; 0.6 V.
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD: Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service personnel. Maintenance performed by unqualified personnel or installation of unauthorized parts could cause personal injury, result in improper equipment performance, invalidate the warranty or result in costly damage.
Page 104
Tabletop Tabletop Extend table Extend Tabletop top 3/16” 3/16" (4.8 mm) Beyond Foot beyond the End of Tilt Frame foot end of the tilt frame. Tilt Frame Tilt frame Figure 6-1. Table Sensor Calibration 8. Display shows ADJUST MODE. 13. Press Reverse Orientation button to save setting. NOTE: During table sensor calibration, note the 14.
22. Display shows BACK DOWN MAX. Lower back NOTE: During table sensor calibration, note the section to maximum position and press Reverse following: Orientation button to save setting. 1) The Reverse Patient LED must be OFF. 23. Display shows BACK UP MAX. Raise back 2) Do not press LEVEL button during procedure section to maximum position and press Reverse or control exits Calibration Mode without...
Page 106
Tabletop Side Member Inner Face Tilt Frame Tilt frame Extend Tabletop 3/16" (4.8 mm) NOTE: Tabletop removed for clarity only. Beyond Foot End Extend of Tilt Frame Figure 6-2. Table Sensor Calibration 26. Press Reverse Orientation button to save setting. 38.
6.4 ACCESSORY ATTACHMENTS 3. Measure lash between inner and outer column tubes. Lash must be no more than 5/64" (2 mm). 6.4.1 Mayo Stand 4. Lower lift column until a small gap exists between outer tube and base plate (see Figure 6-3). Insufficient clearance between table base and floor to allow use of flat-bladed Mayo Stands.
Table 6-1. Articulation Times Tabletop Moved in This Direction Articulation Times (Seconds) Loose Screws Articulation Minimum Target Maximum Trendelenburg 15.0 17.1 19.3 / Reverse Trendelenburg Loose Screws Tilt Right / 10.0 11.4 12.9 Left Back Up / 21.8 24.0 26.2 Tabletop Moved in This Direction Down Kidney Up...
Page 109
Port P Port T Y11 – Kidney Up Y12 – Kidney Down Y21 – Back Up Y22 – Back Down Y32 – Leg Down Y31 – Leg Up Y42 – Height Down Y41 – Height Up Y52 – Rev Trend Y51 –...
WARNING – EXPLOSION HAZARD: Table must not be used in the presence of flammable anesthetics. WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD: • Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service personnel. Maintenance performed by unqualified personnel or installation of unauthorized parts could cause personal injury, result in improper equipment performance, invalidate the warranty, or result in costly damage.
ASSEMBLY have been performed there for several hours and none occur during servicing. To gain access to most of the STERIS 4085 General Try to work in a place with no nearby doors through Surgical Table hydraulic and electrical components, which people are continually passing.
7.4 TABLE BASE COVER ASSEMBLY NOTE: Remove all old sealant and clean all mating parts To gain access to most of the STERIS 4085 General before reapplying the silicone sealant. Use RTV 123 Black Surgical Table hydraulic and electrical components, Silicone Sealant (R005300-568).
Page 113
Figure 7-5. Ensure Base Cover Sits Flush on Figure 7-8. Seal Cover Around Power Inlet Base at Insert Opening 6. Seal around entire perimeter between upper plastic 3. Seal mating areas between front and rear base column covers and column ensuring to fill entire covers as shown in Figure 7-6.
11. Secure lower cover to base using four existing 7. After applying RTV, remove any excess sealant fasteners. using an alcohol dampened rag. 12. Secure base cover to base using screw and spacer. 8. Allow sufficient cure time as recommended by the 13.
6. Carefully slide power supply away from table NOTE: For swivel foot, use thin wrench (P764330-220) frame and gently lift to remove from table taking to carefully remove foot. care not to damage LED display. 8. Install new hydraulic screw and sealing washers to 7.
Note and record various connections for ease of reassembly. Discard all hydraulic screws and sealing washers. NOTE: Refer to 4085 Surgical Table Hydraulics figure in the Illustrated Parts Breakdown section for hose identifications. 7. Remove both stainless-steel socket head cap screws...
3. Turn Main Power Switch to OFF and remove 2. Remove table column cover per S 7.3, T ECTION ABLE battery fuse (refer to supplied Operator Manual for OLUMN OVER SSEMBLY more information). 3. Remove appropriate base cover per S 7.4, ECTION 4.
7. Disconnect P6 and P75 (height sensor) connectors from CPU Board. Tilt 8. Remove both socket head cap screws fastening Cylinder Upper Fastening 6 upp height sensor to table base. Screws 9. Replace sensor and cable by reversing procedure. Connection P75 Connector Lower Panel...
manifold threads and hose connectors are free of rubber seal particles. Column 2) Torque new screws to 12 N•m (106 lbf•in). Cover Cap 3) Ensure distance between cylinder shaft and nut is Upper Connections 1-1/8" (29 mm) - see Figure 7-22. 4) After reassembly, ensure all table functions operate properly.
8. Unscrew Trendelenburg cylinder rod. 7.19 MAIN VALVE BLOCK REPLACEMENT 9. Remove three upper screws from Trendelenburg cylinder side column plastic cap and remove cap. Replace the main hydraulic valve block as follows: NOTE: If bellows are damaged, remove and replace at 1.
4) Tighten bolt from beneath table before returning cylinder mounting plate. 5) Hand tighten all bolts first. 6) Follow 4085 Surgical Table Hydraulics figure in the IPB Section of this manual to ensure proper reconnection. 7) Ensure all screws are secured with Loctite 242.
7.22 KIDNEY CYLINDER REPLACEMENT Replace kidney hydraulic cylinder as follows: 1. Position tabletop to level. 2. Turn Main Power Switch to OFF and remove battery fuse (refer to supplied Operator Manual for more information). 3. Remove tabletop seat section per S 7.20.1, ECTION 5) After reassembly, move kidney section up and down...
6. Remove seat section accessory rail assembly by removing and saving three screws and spacers. 7. Remove top cover. 8. Remove wire protection cover. 9. Disconnect switch from main CPU board (P77 connector). Switch 10. Remove both screws fastening switch to frame and replace switch (see Figure 7-43).
5. Remove Trendelenburg sensor and disconnect 7.30 SLIDE (TRANSLATION) SENSOR from connector (see Figure 7-44). REPLACEMENT 6. Reassemble in reverse order and adjust tabletop to Replace slide (translation) sensor as follows: level. 1. Position tabletop to level, slide to maximum head end.
4. Remove seat plastic cover as follows: 3. Remove tabletop seat section per S 7.20.1, ECTION ECTION a. Remove four top and four side screws from both head-side and foot-side seat cover 4. Remove seat plastic cover as follows: (see Figure 7-33). a.
7.33 BACK SECTION HYDRAULIC 12. Loosen bottom shaft locking screw and remove bottom shaft (see Figure 7-53). CYLINDER REPLACEMENT Replace back section hydraulic cylinder as follows: 1. Position tabletop to level. 2. Remove tabletop seat section per S 7.20.1, ECTION ECTION 3.
2. Turn Main Power Switch to OFF and remove 2. Remove appropriate base cover per S 7.4, ECTION battery fuse (refer to supplied Operator Manual for ABLE OVER SSEMBLY more information). 3. Unlock table and lower tabletop to lowest position. 3.
Page 130
seat section side member to the stationary tilt frame. A small CAUTION – POSSIBLE EQUIPMENT gap exists between the sliding section and the tilt frame. This DAMAGE: Do not remove the bearing gap may be filled with a shim during bearing assembly cartridges from the bearing rail.
Page 132
Table 1 provides part numbers and figure references WARNING – PERSONAL INJURY AND/OR for the most commonly replaced items on a STERIS 4085 EQUIPMENT DAMAGE HAZARD: Repairs General Surgical Table and Table 2 does the same for and adjustments to this equipment must be the hydraulics assembly.
Page 133
FIG. & ITEM UNITS PER PART DESCRIPTION ASSEMBLY NUMBER 10-1 P 146667 035 Harmony LA Wall Control Assembly, Domestic......X P 146667 327 Harmony LA Wall Control Assembly, Global........P 146667 358 Harmony LA Wall Control Assembly, LA2 ........P 146667 278 ASSEMBLY, Cover/Membrane Replacement ........
Page 135
Power Supply, 120V 542200-902 Figure 4 Power Supply, 220V 542201-902 Figure 4 Electric Unit (PC Board) P150832-769 Figure 4 4085 Hand Control Assembly (Without Cable) P150832-632 Figure 6 4085 Hand Control Cable P150832-632 Figure 1 Override Hand Control Assembly 413724-403 Figure 6...
Page 136
Table 2. Quick Reference Parts (Hydraulics) 4085 Surgical Table Reference Hydraulic Assembly Trendelenburg, Cylinder Assembly With Hoses P755718-040 Figure 5 Tilt, Cylinder Assembly With Hoses P755718-041 Figure 5 Height, Cylinder Assembly With Hoses P755718-042 Figure 4 and 5 Leg Right, Cylinder Assembly With Hoses...
Page 137
Table 3. Quick Reference Parts (Labels) Type Part Description Part Number Nomenclature/Location Label Label, Base Warning 413724-362 “Warning - Base Storage” Label Label, Base Warning 413724-361 Size: 11-5/16 x 5-5/32 Base: Head End Label Label, Base Warning 413724-350 Hand/Foot Control, E-Serve Label Label, Leg Position 542200-344...
Page 138
Table 4. Quick Reference Hardware Listing Unit Of Quantity per Type Number Description Measure Unit Cable Tie 084104-001 Cable Tie, 4 129360-598 Cable Tie Clamp 605130-003 Cable Clamp, 3M DIA 129361-169 Cable Clamp, Alum, 5/16 OD M x #8 Std 150830-770 Nylon Cable Clamp, 1/4 ID 150830-771...
Page 139
Table 4. Quick Reference Hardware Listing (Cont’d) Unit Of Quantity per Type Number Description Measure Unit Screw (Cont’d) 600530-425 Socket Head Cap Screw, M4 x 25 SS 600530-508 Socket Head Cap Screw, M5 x 8 SS 600530-510 Socket Head Cap Screw, M5 x 10 SS 600530-512 Socket Head Cap Screw, M5 x 12 SS 600530-608...
Page 140
Table 4. Quick Reference Hardware Listing (Cont’d) Unit Of Quantity per Type Number Description Measure Unit Sealant 764322-636 Hydraulic Oil Pack EA (1 GAL) 005300-545 Loctite * Sealant #222-31 ® 005300-548 Loctite 271-21 Adhesive 005300-554 Thread Locker - 242 EA (50ML) 005300-563 RTV-108 Clear Adhesive EA (2.8 oz)
Page 141
Patient Head Patient Left Patient Right Patient Foot View at A NOTE: Not all labels are shown. See Quick Reference List for part numbers. Figure 1. 4085 Surgical Table Plastic Shroud and Support Assembly (1 of 2) 764333-767...
Page 142
Patient Head Patient Left 16 15 Patient Right Patient Foot View at A NOTE: Not all labels are shown. See Quick Reference List for part numbers. Figure 1. 4085 Surgical Table Stainless-Steel Shroud and Support Assembly (2 of 2) 764333-767...
Page 143
UNITS PER FIG. & PART DESCRIPTION ASSEMBLY ITEM NUMBER 4085 Surgical Table Plastic Shroud and Support Assembly ..4085 Surgical Table Stainless-Steel Shroud and Support Assembly ..................P 150832 647 BASE, Cover ..................– P 150832 520 COVER, Base, Short, SS (Head) ............
Page 144
Patient Left Figure 2. 4085 Surgical Tabletop (1 of 3) 764333-767...
Page 145
Back Section Headrest Section Seat Section Leg Section Figure 2. 4085 Surgical Tabletop (2 of 3) 764333-767...
Page 146
Typical of Rail Assemblies See Figure 7 Patient Head Back Section Members, Patient Foot Not Covers Figure 2. 4085 Surgical Tabletop (3 of 3) 764333-767...
Page 148
UNITS PER FIG. & PART DESCRIPTION ASSEMBLY ITEM NUMBER 4085 Surgical Tabletop ..............P 755717 727 KIT, Leg End- Right Covers with Screws ..........P 755717 728 KIT, Leg End-Left Covers with Screws..........542200 344 • LABEL, Leg Position .................
Page 149
Detail A See Detail A See Detail C See Detail E See Detail D Detail B Patient Right Detail C 10 11 12 Detail E Detail D Figure 3. 4085 Surgical Tabletop Slide and Section Assembly (1 of 4) 764333-767...
Page 150
See Detail F See Sheet 4 of 4 Patient Right Patient Left Detail F Slide Motor Figure 3. 4085 Surgical Tabletop Slide and Section Assembly (2 of 4) 764333-767...
Page 151
Patient Left See Sheet 4 of 4 See Figure 8 Figure 3. 4085 Surgical Tabletop Slide and Section Assembly (3 of 4) 764333-767...
Page 152
Figure 3. 4085 Surgical Tabletop Slide and Section Assembly (4 of 4) 764333-767...
Page 153
UNITS PER FIG. & PART DESCRIPTION ASSEMBLY ITEM NUMBER 4085 Surgical Tabletop Slide and Section Assemblies ....342273 555 ASSEMBLY, Leg Section, Locking Device........... 542200 029 • KEY, Lock ..................632500 055 • SPRING, Compression, 8 x 16mm ............ 542200 221 ASSEMBLY, Cable, Switch/Leg Section ..........
Page 154
PART DESCRIPTION ASSEMBLY ITEM NUMBER 4085 Surgical Tabletop Slide and Section Assemblies....HEAD AND LEG SECTIONS ASSEMBLY, Leg Section (See Figure 8)..........TOP SECTION AND RAILS (See Figure 2; Headrest, See Figure 7) COVERS, LEG SECTION (See Figure 2) 542200 270 PROTECTOR, Articulator, #5 ..............
Page 155
4085 Surgical Tabletop Slide and Section Assemblies ....600620 505 SETSCREW ..................542200 088 ROD, Center..................542200 087 ROD, Back Section ................* Order 4085 Hydraulic Screw Parts Pack (342291-555) for replacement hydraulic screws and sealing washers. ** Individual part number not available. 764333-767...
Page 156
23 24 Figure 4. 4085 Surgical Table Base Assembly (1 of 2) 764333-767...
Page 157
Figure 4. 4085 Surgical Table Base Assembly (2 of 2) 764333-767...
Page 158
PART DESCRIPTION ASSEMBLY ITEM NUMBER 4085 Table Base Assemby With Plastic Shrouds ......4085 Table Base Assemby With SS Shrouds ........ WELDMENT, Base (Not Service Component) ........P 764332 718 PACK, Casters (Set of Four) ..............A/R A/R P 150832 207 •...
Page 159
UNITS PER FIG. & PART DESCRIPTION ASSEMBLY ITEM NUMBER 4085 Table Base Assemby With Plastic Shrouds ......4085 Table Base Assemby With SS Shrouds......... P 150832 302 SPACER, Floor Lock Cylinder ............. 602321 262 WASHER, Flat, M5 ................P 150832 515 ASSEMBLY, Auxiliary Pedal Weld............
Page 160
PART DESCRIPTION ASSEMBLY ITEM NUMBER 4085 Table Base Assemby With Plastic Shrouds ......4085 Table Base Assemby With SS Shrouds ........ 342213 555 SET, Fuse (Not Shown) ............... • FUSE, Main 6, 3A T (F6/7) ..............• FUSE, Battery, 16A (F8) ..............
Page 161
See 2 of 2 Figure 5. 4085 Surgical Table Hydraulics (1 of 2 ) 764333-767...
Page 165
† VOLTAGE SUPPRESSOR Sliding Motor Assembly P 150832 FOOT CONTROL, 4085 Table (Not Shown) * Item #23 is not included with Item #19. † Item #24 is not included with sliding motor assembly. Item #8 is not included in Pump 342211-555.
Page 167
FIG. & PART DESCRIPTION ASSEMBLY ITEM NUMBER 150832 559 4085 Surgical Table Headrest Aassembly........FRAME, Headrest LH, 4085 ..............FRAME, headrest, RH, 4085 ............... P 150823 276 SPACER, Siderail, 9/16 ............... P 150832 379 CAP PIVOT, Chrome Painted.............. P 056397 899 SIDE RAIL, Bent Headrest, LH ............
Page 168
Patient Right Figure 8. 4085 Surgical Table Leg Section Assembly 764333-767...
Page 169
Desc. Item # Number 413724-543 LEG SECT ASSY Cmax/4085/5085 General P150832-760 LEG SECT ASSY, Featherweight Cmax/4085/5085 General (Not Shown) ASSEMBLY, RT LEG SECTION. MEMBER ASSEMBLY, LEFT LEG SECTION MEMBER 3 624900-014 BRING,SLEEVE 14 MM 4 624900-015 BEARING, SLEEVE 8MM # 6...
Page 170
Front Cover Cover) changes/manual errors and clarifications A: Stamped to show Live Document Date P: Wrong figure number shown 4/ 4085 Table Base Assembly C: Typo A: Should be figure 4 P: Wrong figure number shown 7/ 4085 Surgical Table...
Page 171
These drawings are supplied as an aid in understanding STERIS 4085 General Surgical Table operation and troubleshooting. See Table 1 for listing and order of drawings. Table 1. STERIS 4085 General Surgical Table Technical Drawings Title Sheet Drawing Reference Number...
Need help?
Do you have a question about the 4085 and is the answer not in the manual?
Questions and answers