PM100USB

9 Appendix

9.1 Technical Data

General Data
Detector Compatibility
Display Type
Display Update Rate (max)
Display Format
Current Input (Photodiode Sensors)
Connector
Units
Measurement Ranges
Display Resolution
AD Converter
Measurement Uncertainty
Input Bandwidth (Analog)
Wavelength Correction
Beam Diameter Setting
Voltage Input (Thermopile Sensors)
Connector
Units
Measurement Ranges
Display Resolution
AD Converter
Measurement Uncertainty
Input Bandwidth (Analog)
Wavelength Correction
Beam Diameter Setting
32
Photodiode Sensors S1xxC Series
Thermal Sensors S3xxC / S4xxC Series
Pyroelectric Sensors EC1xxC/ES2xxC Series
Photodiodes (max. 5 mA)
Thermopiles (max. 1 V)
External PC, Laptop, Tablet - Windows Application
Up to 300 Hz, depending on PC and Settings
Optical Power Monitor:
Numerical, Tuning Needle, Bar Graph, Trend Graph, Statistics, His-
DE-9 female, left side
Ranges selectable in W or A, Sensor Dependent
1 pA / responsivity value (A/W)
± 0.2 % f.s. (5µA - 5mA)
DC - 100kHz, depending on sensor and settings
DE-9 female, left side
Ranges selectable in W or V, Sensor Dependent
1 µV / responsivity value (V/W)
± 0.5 % f.s. (10 mV - 1 V range)
± 1.0 % f.s. (1 mV range)
DC - 10Hz, depending on sensor and settings
Sensor depending; nm, (V/W)
Pyros (max. 100 V)
togram
W, dBm, W/cm², A
6
10
; 50 nA - 5 mA
16 bit
± 0.5 % f.s. (50nA)
nm (A/W)
1/e²
W, dBm, W/cm², V
4
10
; 1 mV - 1V
16 bit
1/e²