PM100USB
7.1.2.3 PM100USB specific SCPI Command Reference
See also SCPI Specification, Version 1999.0, May, 1999, http://www.scpiconsortium.org. All
commands with a 'SCPI' checkmark are described in the SCPI specification.
All described commands work with the PM100D, PM100A, PM100USB and PM200 instruments
(with some limitations due to the hardware capabilities).
7.1.2.3.1 SYSTem subsystem commands
Command
SYSTem
:ERRor
[:NEXT]?
:VERSion?
:LFRequency <numeric val-
ue>
:LFRequency?
:SENSor
:IDN?
7.1.2.3.2 STATus subsystem commands
Command
STATus
:MEASurement
[:EVENt]?
:CONDition?
:PTRansition <value>
:PTRansition?
:NTRansition <value>
:NTRansition?
24
Path to SYSTem subsystem.
(SCPI Vol.2 §21)
Returns the latest <error code, "message">.
(SCPI Vol.2 §21.8.8)
Query level of SCPI standard (1999.0) .
(SCPI Vol.2 §21.21)
Sets the instrument's line frequency setting to 50 or 60Hz.
(SCPI Vol.2 §21.13)
Query the instrument's line frequency setting.
(SCPI Vol.2 §21.13)
Query information about the connected sensor. This is a query
only command. The response consists of the following fields:
<
>,<
>,<
name
sn
cal_msg
<name>
Sensor name in string response format
<sn>
Sensor serial number in string response format
<cal_msg>
Calibration message in string response format
<type>
Sensor type in NR1 format
<subtype>
Sensor subtype in NR1 format
<flags>
Sensor flags as bitmap in NR1 format.
Flag:
Is power sensor
Is energy sensor
Response settable
Wavelength settable
Tau settable
Has temperature sensor
Path to STATus subsystem. (SCPI Vol.2 §20)
Path to control measurement event registers
Read the event register
Read the condition register
Program the positive transition filter
Read the positive transition filter
Program the negative transition filter
Read the negative transition filter
Description
>,<
>,<
>,<
type
subtype
Description
>
flags
Dec.value:
1
2
16
32
64
256
© 2020 Thorlabs