7.2 OON ("Out Of Neutral")

OON ("Out Of Neutral") is a safety feature that prevents
accidental operation of mobility device functions when the
system's primary input is in an out of neutral position.
For proportional joysticks, an out of neutral position is when
the joystick is outside or greater than the neutral window.
For discrete (switch) joysticks, an out of neutral position
is when the joystick is outside or greater than the switch
threshold. For switches, out of neutral is when one or more
switches are activated.
An OON indication is displayed when the primary input is
out of neutral and one of the following occurs:
the system is powering up,
after a function change,
when the system comes out of an inhibit or drive
lock-out,
when the settings menu is exited,
when indirect navigation is exited,
when Rest is exited or
on Live Handover.
1637423-H
OON activation is slightly different between a drive
function and a non-drive function for quadrants that
have no programmed output.
For non-drive functions, such as seating, OON
is not activated if the primary input is in an
out of neutral position in a quadrant, that has
not been programmed for an output. This is
particularly useful, for example, for Head Array
users who may have just the left and right pads
programmed for output, allowing to rest the
head on the middle pad without activating an
OON.
For drive functions, no matter how the
quadrants are programmed, OON is always
activated when the primary input is out of
neutral when powering up or when waking
from sleep.
Drive OON Warning
Troubleshooting
Fig. 7-1
121