11. Press the "Enter" button to set the O
automatically.
12. If the oxygen sensor has been replaced, fill out the
values of the oxygen sensor constants of the calibration
certificate. Do not change the constants if the oxygen
sensor will be calibrated.
Use the arrow buttons to enter the digits and the "Enter"
button to move to the next digit. After the last digit,
press the "Enter" button to proceed to the next menu
point. Select the menu point "Previous Menu" to return
to the calibration menu.
13. Press the "Enter" button to save the calibration. If the
calibration was successful, the c-DGM software
program will return to the main menu. If an error has
occurred, you will receive the notification:
"Questionable calibration, old calibration
restored"
14. You can also exit the "Calibration" menu without
saving the calibration. Select the menu point "Main
menu" and press "Enter" to do so.
c-DGM/i-DGM
sensor constants
2
, &&D E^ 'D 'D
CALIBRATION
P-ATM.
:1.013 BAR
ZERO PNT : 0.000 %
O
HIGH
: 4.010 %
2
O O O O
SENSOR CONSTANTS
SENSOR CONSTANTS
SENSOR CONSTANTS
SENSOR CONSTANTS
2 2 2 2
STORE CALIBRATION
MAIN MENU
O2 SENSOR CONSTANTS
C1: + 0.86469
C2: - 0.03803
C3: + 0.00852
C4: + 28.98344
PREVIOUS MENU
CALIBRATION
P-ATM.
:1.013 BAR
ZERO PNT : 0.000 %
O
HIGH
: 4.010 %
2
O
SENSOR CONSTANTS
2
STORE CALIBRATION
STORE CALIBRATION
STORE CALIBRATION
STORE CALIBRATION
MAIN MENU
CALIBRATION
P-ATM.
:1.013 BAR
ZERO PNT : 0.000 %
O
HIGH
: 4.010 %
2
O
SENSOR CONSTANTS
2
STORE CALIBRATION
MAIN MENU
MAIN MENU
MAIN MENU
MAIN MENU