Installing the Wall-Mount
(cont.)

Wiring the Wall-Mount

2-4
4X
Ø0.38
[Ø10]
Figure 2-5: OptiSonde Wall-Mount - Mounting Hole Locations
For mounting, secure the unit to a wall or panel using all four
mounting holes. Use stainless hardware, a minimum of 2 inches long,
#8 screws with washers. Screws should go directly into wood studs.
If no stud is available, use suitable hardware for the application, such
as with plasterboard or cement.
The wall-mount unit meets IP65 rating for water ingress
when tested with solid plugs. In order to maintain this
rating, installation personnel must ensure that each of the
two wiring ports use a single cable with a minimum
diameter of 0.24 in (6 mm) and a maximum diameter of
0.47 in (12 mm). Recommended wire nuts torque is 37 in-
lbs (4.2 Nm). Any unused wiring port shall be plugged with
2.86
[73]
Note: units are inches [mm]
Caution!
a supplied solid plug.