Power Off/ Shutdown - GE LOGIQ F Series Basic Service Manual

Hide thumbs Also See for LOGIQ F Series:
Table of Contents

Advertisement

GE
D
5446617-100, R
IRECTION
4-3-2

Power Off/ Shutdown

NOTE:
After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
4-3-2-1
Back-end Processor Power Down
To power down the system:
1.) Press the Power On/Off switch at the front of the system once.
2.) The System-Exit window is displayed.
3.) Using the Trackball or Select key, select Shutdown.
4.) The shutdown process takes a few seconds and the power off sequence is complete when the
power status LED is turned amber.
5.) Disconnect the probes.Clean or disinfect all probes as necessary. Store them in their shipping
cases to avoid damage.
4-3-2-2
Scanner Shutdown
Disconnect the Mains Power Cable is necessary. For example: Relocating the scanner.
CAUTION DO NOT unplug and/or transport the unit until after the power off sequence has been completed.