Binaural Screening; Preparing For A Binaural Screening; Binaural Screening Result - Maico MB 11 Classic Operation Manual

Hide thumbs Also See for MB 11 Classic:
Table of Contents

Advertisement

Operation Manual MB 11 Classic

6.5 Binaural Screening

With the MB11 Classic, simultaneous binaural testing is possible. This allows
both ears to be screened at the same time. Testing will continue until both
ears achieve a Pass result or until one or both ears' test terminates in a Refer
result. To perform a "binaural" test, the setting "Binaural Test Allowed" in
the Options dialog must be enabled. Otherwise the selection of Binaural in
the ear button control will not be possible.

6.5.1 Preparing for a Binaural Screening

Prepare the patient as described in chapter 6 but apply the disposable ear
tips or EarCups and insert earphones for both ears.
Select the "Ear" button on the Main window or the "Other Ear" on the
measuring screen until "Binaural" (in pink letters) is displayed. Click on
"Start Measurement" to begin measurement. The stimulus and the data
collection will begin for both ears simultaneously.

6.5.2 Binaural Screening Result

Unlike monaural test, the binaural measurement will display two colored
lines, red for the right ear and blue for the left. Both ears are screened for
hearing loss at 35dBnHL. If responses are detected from both ears, the lines
move upward until it reaches the 100% area colored green, then the result
will be shown as "PASS".
Figure 23
MB 11 – Binaural screening program Test result "PASS"
8100498-4_OM_MB11_Classic_3.18_EN_16a.docx
45
01/16

Advertisement

Table of Contents
loading

Table of Contents