Operation Manual MB 11 Classic
6.4.5 Test result "Refer"
If 100% of the pass criteria is not reached after 180 seconds of data
Figure 21
MB 11–program Test result "REFER"
collection time, the test is automatically terminated and a "REFER" will be
displayed below and to the right of the graph.
In this case, follow your screening program protocol to determine whether
another screening should be performed or whether the baby will be referred
for a diagnostic follow-up test.
When you leave the test screen by clicking on the button "QUIT", the test
result will be stored automatically in the data base.
8100498-4_OM_MB11_Classic_3.18_EN_16a.docx
43
01/16