Integrating the i.Profiler
network
• Connect the i.Profiler
cable.
• Switch the i.Profiler
software by pressing the Close Application button in the Shut Down
window and switch to the Microsoft Windows desktop (administrator
rights required).
• Double-click the Network Configurator symbol on the desktop. The
Network Configuration Tool dialog appears.
• Select the Ethernet connection used by your network in the Network
Adapter field. The left connection is marked [00000000], the right
connection [00000001].
• Enter the new IP address in the IP Address field. The three first digit
blocks must conform to the digit blocks of your network. The last digit
block can be selected, but the number must be unique.
• Change the subnet mask in the Subnet Mask field.
• Click on the Change IP button. The message "The static IP was set
successfully." appears.
• If available, you should enter Gateway and DNS-Server.
• Now you can close the Network Configurator.
• Double-click on the Save Configurations program icon on the
Microsoft Windows desktop to save the new configuration. The device
will restart.
® plus
into an existing local
® plus
to the practice network (LAN) using a network
® plus
on. After the start, quit the i.Profiler
Installation
® plus
000000-1520-928_GA_EN_021013