Pavone Systems DAT 1400 Technical Manual page 51

Weight transmitter/indicator with rs232 serial, analog and fieldbus output
Table of Contents

Advertisement

0-trAC
ZERO TRACKING [1306]
This function allows you to perform temporary zero calibration compensating for the temperature drift
of the weight.
Switching off the transmitter automatically restores the previous zero calibration.
The maximum weight that can be reset by this parameter is 2% of the capacity of the system.
To disable this function, set the value 0.
Value
Variation
0
Control excluded
1
0.5 div/sec
2
1 div/sec
3
2 div/sec
4
3 div/sec
Default: 0
0-bAnd
ZERO BAND [1307]
This parameter defines the number of divisions that can be reset by pressing the zero button on the
front or the associated Input.
Values: from 0 to 200.
Default: 100
Page 49

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DAT 1400 and is the answer not in the manual?

Table of Contents