SERIAL COMMUNICATIONS PARAMETERS
This menu makes it possible to configure the COM1 and COM2 serial ports and the communication parame-
ters. The instrument has two independent serial ports:
COM1 always with interface RS232 ; COM2 can be fitted with either of the following interfaces:
RS485, ETHERCAT, ETHERNET, ETHERNET IP, PROFINET.
C1 Nod. .
RS232 OUPUT MODE
Selecting the value transmitted on output RS 232.
Values that can be selected:
nEt
GroSS
PEAk
Default: nEt
C1Prot. . .
COM1 PROTOCOL
It defines how to use the RS232 serial port:
Values that can be selected:
None: Serial communication OFF
Contin: Continuous transmission of the weight string. It can be used, for example, to drive a weight
repeater. See details in the relevant section.
on deM: When the Operator presses the relative button on the front or uses Input 2, a weight string
is sent. The command is acceptedif the weight is stable. Between two consecutive transmissions,
variation of the weight must be at least equal to 20 divisions.
Autom: A weight string is sent automatically when theweight stabilises at a value above the minimum
weight (20 divisions).
Slave: ASCII protocol. See details in the relevant section.
Print: When the operator presses the key on the front or by Input, a weight string is transferred. The
command is accepted if the weight is stable. Between two successive transmissions, the weight must
have a variation of at least 20 divisions.
Default: Slave
Page 35