PRINTER PROTOCOL
Data transmission protocol to Plus Printer
Printing can be started by pressing a key (see section FUNCTION OPERATIONAL) or by input (see
paragraph SETTING I/O).
Here is an example of printer.
Date is printed only in case of hardware with time clock.
The peak value is only printed if the peak function is enabled.
The identification code of the weighing is printed only in case of METRIC operation and
alibi memory configured.
The conditions to printing are:
• Stable weight (or stabilized within 3 seconds from command).
• Since the last executed weigh, the weight has undergone a change of at least 20 divisions (delta
weight).
• Gross weight equal to or greater than the minimum weight (20 divisions) and less than the maxi-
mum capacity.
• Net weight not null.
Only in case of FREE functioning, printing is permitted even with gross weight less than the minimum
weight or with zero net weight.
216/06/16
15:32
Net
209.0 kg
Gross
211.5 kg
Tare
2.5 kg
Peak
268.5 kg
Code
212456
Page 63