24 Annex
Electromagentic immunity, IEC 60601-1-2: 2015, based on table 4, 5
Immunity test
IEC 60601
test level
Conducted RF
3 V
eff
IEC 61000-4-6
150 KHz to
80 Mhz
6 V in ISM-
and amateur radio
frequency bands
between 150 kHz
and 80 MHz
80% AM to 1 kHz
3 V/m
80 MHz to 2,7 GHz
Radiated RF
IEC 61000-4-3
80% AM at 1 kHz
Note 1:
At 80 Hz and 800 MHz, the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radios, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, and electromagnetic site survey should be considered. If the measured field strength in the
location in which the Myopia Master® is used exceeds the applicable RF compliance level above, the Myopia Master® should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relo-
cating the Myopia Master®.
b. Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
86 / 92
Compliance level
Electromagnetic environment - Guidelines
Portable and mobile RF communications equip-
ment should be used no closer to any part of
Myopia Master®, including cables, than the re-
commended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d
=
d
=
V
= 3 V
eff
d
=
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Field strength from fixed RF transmitters, as
determined by an electromagnetic site survey
(a), should be less than the compliance level in
each frequency range (b).
Interferences may occur in the vicinity of equip-
ment marked with the following symbol:
Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
3 5 ,
- - - - - - - - -
P
V 1
3 5 ,
- - - - - - - - -
P
80MHz to 800 MHz
E 1
7
- - - - - - - - -
P
800 MHz to 2.5 GHz
E 1