Ports (Esd Nx30, Nsg 438 Models Only) - Ametek esd NX30 Quick Start Manual And Safety Instructions

Table of Contents

Advertisement

AMETEK CTS
5.1.2

Ports (esd NX30, NSG 438 models only)

This port is indented to allow the user to connect external signals in order to remote control the generator, as well as to con-
nect external accessories like the charge removing device. See tables and graphs below for detailed signal description and
drive circuitry information.
Connector J3: Pin Assignment
Pin
Signal Name
1
EXT_TRIGGER
2
NC
3
NC
4
EUT_FAIL
5
GND
6
NC
7
Charge remove
8
GND
9
+15V
Ext. Trigger (esd NX30 only):
The following circuit is built in behind the external trigger input connector. This function is similar to the function of the yellow
trigger knob on the handle. Trigger signal active low, i.e. to be pulled to ground.
Charge removing connection:
This function is to drive an external bleedoff switch option or an external relay.
Quick Start and Safety Guide
Description
External trigger input
-
-
EUT failure input
(reserved for future use)
Earth
-
Charge remover drive output
Earth
Voltage output (max 500 mA)
V 1.0
ESD Simulators
13 / 17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Esd nx30.1DitoNsg 435Nsg 437Nsg 438Nsg 438a

Table of Contents