Two wire point level switch auto-calibration or manual calibration selectable

Advertisement

Two Wire Point Level Switch
Auto-Calibration or Manual Calibration
Telephone:
Fax:
E-mail:
drexelbrook.service@ametek.com
Website:

Operating Instructions

ThePoint
Selectable
+1 215-674-1234
+1 215-674-2731
www.drexelbrook.com
Level Measurement

Installation and

Series
TM
Leader in

Advertisement

Table of Contents
loading

Summary of Contents for Ametek ThePoint Series

  • Page 1: Installation And

    Leader in Level Measurement Installation and Operating Instructions ThePoint Series Two Wire Point Level Switch Auto-Calibration or Manual Calibration Selectable Telephone: +1 215-674-1234 Fax: +1 215-674-2731 E-mail: drexelbrook.service@ametek.com Website: www.drexelbrook.com...
  • Page 2 AMETEK Drexelbrook makes no warranty of any kind with regard to the material contained in this manual, including, but not limited to, implied warranties or fitness for a particular purpose. Drexelbrook shall not be liable for errors contained herein or for incidental or consequential damages in connection with the performance or use of material.
  • Page 3 EDO # 05-19-110 PMT1-LM Issue# 13 ThePoint Series Two Wire Point Level Switch Auto-Calibration or Manual Calibration Selectable...
  • Page 4 Contents...
  • Page 5: Table Of Contents

    Contents Section 1: Introduction ......................1 System Description ....................1 Technology ......................1 Model Number ...................... 2 Housing Dimensions ................... 4 Section 2: Installation ......................5 Unpacking ......................5 Mounting and Installation Guidelines ..............5 Input Wiring ......................8 Output and LED Status ..................8 Electronic Unit .....................
  • Page 6 Section 1...
  • Page 7: Section 1: Introduction

    Cote-Shield™ circuitry within the Electronics Unit. This patented Cote-Shield™ circuitry is designed into Figure 1-1 ThePoint series and enables the instrument to ignore the Simple Capacitance effect of buildup or material coating on the sensing element. Probe...
  • Page 8: Model Number

    The Point™ Series Introduction Model Number Technology RF Admittance Measurement Type Std Auto Cal Hi Sense .5 pF Auto Cal All Calibration modes are built Std 2 pF Fixed Hi Sense .5 pF Fixed into the standard unit. Modes 10 pF Auto Cal Hi Sense Manual can be changed in the field as 10 pF Fixed...
  • Page 9 low pressure 1.4 bar @ 232°C (20 PSI @ 450°F) Heavy duty, agitated 700-0202-043 remote 69 bar @ 38°C (1000 PSI @ 100°F) 316SS and TFE tanks or material 13.8 bar @ 232°C (200 PSI @ 450°F) with high bulk density (1) The Point™...
  • Page 10: Housing Dimensions

    The Point™ Series Introduction Housing Dimensions 4.5" (114mm) 6.5" (165mm) 4.75" (121mm) Figure 1-3 Compartment Housing Detail Equipotential Terminal (CENELEC Systems)
  • Page 11: Section 2: Installation

    ThePoint™ Series Installation Section 2: Installation Unpacking Carefully remove the contents of the shipping carton and check each item against the packing list before destroying any packing material. If there is any shortage or damage, immediately report it to the factory at + 215-674-1234. Mounting and Installation Guidelines WARNING: ThePoint equipment is rated explosion-proof.
  • Page 12 Installation Mounting and Installation Guidelines (continued) After system is installed and level is below sensing element, apply power. ThePoint series instrument does not require The Point RF any calibration or setpoint adjustments and is ready to detect instrument is rated change in level.
  • Page 13 ThePoint™ Series Installation Mounting and Installation Guidelines (continued) • If installation area is rated explosion-proof and requires conduit seal fittings, they should be used in accordance with company standards and local codes. • Mounting sensing element inside a pipe is not recommended.
  • Page 14: Input Wiring

    ThePoint™ Series Installation Input Wiring WARNING: If ThePoint instrument is located in a hazardous environment, do not open enclosure cover or make/break any electrical connections without first disconnecting electrical power at the source. Ensure that wiring, electrical fittings, and conduit connections conform to electrical codes for the specific location and hazard level.
  • Page 15: Electronic Unit

    ThePoint™ Series Installation Electronic Unit Remove housing lid to access status LEDs, time delay adjustment, and configuration jumpers. See Figure 2-5. OUTPUT Right = 8mA Alarm 16 mA Normal GREEN LED - POWER Left = 8mA Normal RED LED - ALARM 16 mA Alarm TIME DELAY JUMPER RESET/RECAL...
  • Page 16 ThePoint™ Series Installation 2.5.4 Failsafe Failsafe describes the level condition that causes the transmitter to go into alarm. • High Level FailSafe (HLFS) is the condition when the probe is covered, the unit goes into alarm. • Low Level FailSafe (LLFS) is the condition when the probe is uncovered, the unit goes into alarm.
  • Page 17: Spark Protection

    ThePoint™ Series Installation Spark Protection Applications involving insulating granulars and insulating liquids may produce a static discharge that can damage the electronics. The RF series instrument is supplied with integral heavy-duty spark protection to prevent static discharges from damaging the electronic circuits. Sensing Element Connection (Integral Sensing Elements) Sensing element connects to the rear side of the circuit board and is factory-installed.
  • Page 18 ThePoint™ Series Installation Sensing Element Connection (Continued) For ThePoint instruments mounted remotely from sensing element, cable connections from sensing element to Electronic Unit are made to terminals beneath the Electronic Unit. See Figure 2-7. 3 TERMINAL PROBE (TYP) COTE-SHIELD EQUIPOTENTIAL (RED) TERMINAL GROUND...
  • Page 19: Calibration

    ThePoint™ Series Installation Calibration ThePoint™ level measurement switch features both Auto- Cal and manual calibration. The standard Auto-Calibration mode is applicable to most liquids and slurries point level measurements. If preferred, the manual calibration can be used and is recommended for some application. ThePoint electronic unit has auto and manual calibration modes built into the standard unit and can be accessed through a simple routine (see section 2.8.5).
  • Page 20 ThePoint™ Series Installation 2.8.1 Selecting the Calibration Mode for your application (Continued) Application Guide (For instructions on how to access alternate modes see 2.8.4) Application Calibration Mode Liquids and Slurries Auto-Cal Mode #2 Granular /Solids with Bulk Density Manual Cal Mode #7 greater than 20#’s per cubic foot Granular/Solids with Bulk Density Manual Cal Mode #8 (high...
  • Page 21: Calibration Procedures

    ThePoint™ Series Installation 2.8.3 Using ThePoint with Manual Calibration modes #7, and 8 Warning! Before removing the explosion-proof housing cover in a potentially hazardous are, make certain that the area is safe. When calibration is complete, the cover must be replaced. Make sure that ThePoint is set to either mode #7 (standard Sensitivity) or mode #8 (high sensitivity).
  • Page 22 ThePoint™ Series Installation 2.8.3 Manual Calibration modes #7, and 8 (Continued) Manual Calibration When material level can be moved Make certain that ThePoint is in manual calibration mode #7 or 8 See Section 2.8.5 Configuration Adjustment RED LED Notes Settings Potentiometer Fail Safe = High Level Turn counter clockwise...
  • Page 23 ThePoint™ Series Installation 2.8.3 Manual Calibration modes #7, and 8 (Continued) Manual Calibration When material level can not be moved Make certain that ThePoint is in manual calibration mode #7 or 8 See Section 2.8.5 Configuration Adjustment RED LED Notes Settings Potentiometer Fail Safe = High Level...
  • Page 24 ThePoint™ Series Installation 2.8.3 Manual Calibration modes #7, and 8 (Continued) Note 1: Most water based materials can be considered conductive, such as acids, bases, salt solutions, water based slurries, and very wet granular materials. Carbon black and powdered metals conduct even without water.
  • Page 25 ThePoint™ Series Installation 2.8.4 Accessing the Calibration Modes 1. On the top side of ThePoint, temporarily remove the shunt from the “Time Delay Selection Jumper” (see Fig. 2-8) and place it on pins 1 & 2 of the 3-pin connector. (Pin 1 is closet to the LEDS) The green LED will go out and the red LED will begin to flash.
  • Page 26 ThePoint™ Series Installation 2.8.4 Accessing the Calibration Modes (Continued) Definition of Modes Mode 2: Fixed Cal 2pF: 2pF differential, set point locked 2pF above starting capacitance Mode 7: Manual calibration standard sensitivity – pots adjusts from 0 to 65pF Mode 8: Manual calibration High sensitivity – pot adjusts from 0 to 27 pF Mode 6: Fixed Cal 0.5pF: 0.5pF differential, set point locked 0.5pF above starting capacitance...
  • Page 27: Section 3: Troubleshooting

    ThePoint™ Series Troubleshooting Section 3: Troubleshooting WARNING: If ThePoint instrument is located in a hazardous environment, do not open enclosure cover or make/break any electrical connections without first disconnecting electrical power at the source. Ensure that wiring, electrical fittings, and conduit connections conform to electrical codes for the specific location and hazard level.
  • Page 28: Testing Electronic Unit

    ThePoint™ Series Installation Testing Electronic Unit Use the following steps to test the electronic unit: 1. Be sure environment is safe before removing lid from housing. 2. Observe failsafe jumper on circuit board on top of electronic unit (shown in Figure 2-5). Move jumper from current setting to alternate setting [HLFS to LLFS or vice versa].
  • Page 29: Over Range

    ThePoint™ Series Troubleshooting Over Range If the Green LED is flashing, the instrument has detected the uncovered sensing element capacitance exceeds the limits of the transmitter. Consult factory instructions. Under Range If the Red LED is flashing, the instrument has detected the sensing element capacitance is too small.
  • Page 30: Testing Remote Cable

    Shorted wires should read 0 ohms Short out two conductors Factory Assistance AMETEK Drexelbrook can answer any questions about ThePoint series instrument. Call Customer Service at 1- 215-674-1234. If you require assistance and attempts to locate the problem have failed: •...
  • Page 31: Field Service

    Drexelbrook engineers and specialists, and provide detailed information on all aspects of level measurement, including theory and practice of instrument operation. For more information write to: AMETEK Drexelbrook, Communications/ Training Group or call 215- 674-1234. 3.10 Equipment Return...
  • Page 32: 3.11 Rf Point Level Troubleshooting Guide

    ThePoint™ Series Troubleshooting 3.11 RF Point Level Troubleshooting Guide Symptom Possible Cause Solution See Section Switch is in alarm and will not Sensor is coated by a conductive material and the Need a sensor with a longer Cote-Shield Section 2.2 clear Cote-Shield™...
  • Page 33: Section 4: Specifications

    ThePoint™ Series Specifications Section 4: Specifications RF/ Capacitance Technology: Dependant Upon Mode Calibration: High and Low level Modes of Operation: 2mm (0.08 inch) Repeatability: conductive liquids less than 1 second Response Time: 0 to 60 seconds forward Time Delay: and reverse acting -40 to 70°C (-40 to 158°F) Ambient Electronics: -40 to 85°C (-40 to 185°F)
  • Page 34 Section 5...
  • Page 35: Section 5: Approvals Available

    ThePoint™ Series Approvals Available Section 5: Approvals Available Equipment Ratings Intrinsically Safe for use in Class I, II, and, III, Division 1, Groups A, B, C, D, E, F and G in accordance with Entity requirements and Control Drawing 420-0004-220-CD; Nonincendive for use in Class I, Division 2, Groups A, B, C, and D;...
  • Page 36 Section 6...
  • Page 37: Section 6: Control Drawings

    ThePoint™ Series Control Drawings Section 6: Control Drawings FM / FMc Control Drawings...
  • Page 38 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 39 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 40 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 41 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 42 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 43 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 44 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 45 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 46 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 47 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 48 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 49 ThePoint™ Series Control Drawings FM / FMc Control Drawings (Continued)
  • Page 50: Atex Control Drawings

    ThePoint™ Series Control Drawings ATEX Control Drawings...
  • Page 51 ThePoint™ Series Control Drawings ATEX Control Drawings (Continued)
  • Page 52 ThePoint™ Series Control Drawings ATEX Control Drawings (Continued) ATEX AND IECEx APPROVED INTEGRAL 9-14-119 ADD IECEx...UPDATE SENSING ELEMENT OPTIONS SEE TITLE BLOCK APP'D DATE ZONE EDO # DESCRIPTION REVISION HISTORY DREXELBROOK 420-0004-221-CD...
  • Page 53 ThePoint™ Series Control Drawings ATEX Control Drawings (Continued) ATEX AND IECEx APPROVED REMOTE 6-16-111 ADD IECEx...UPDATE SENSING ELEMENT OPTIONS SEE TITLE BLOCK APP'D DATE ZONE EDO # DESCRIPTION REVISION HISTORY DREXELBROOK 420-0004-221-CD...
  • Page 54 ThePoint™ Series Control Drawings ATEX Control Drawings (Continued) ATEX AND IECEx APPROVED REMOTE 6-16-111 ADD IECEx...UPDATE SENSING ELEMENT OPTIONS SEE TITLE BLOCK APP'D DATE ZONE EDO # DESCRIPTION REVISION HISTORY DREXELBROOK 420-0004-221-CD...
  • Page 55 ThePoint™ Series Control Drawings...
  • Page 56: Heavy Duty Spark Protection

    ThePoint™ Series Control Drawings Heavy Duty Spark Protection...
  • Page 57 ThePoint™ Series Control Drawings Heavy Duty Spark Protection (Continued)
  • Page 58: Adding A Padded Capacitor

    ThePoint™ Series Control Drawings Adding a Padded Capacitor...
  • Page 59 ThePoint™ Series Control Drawings Adding a Padded Capacitor (Continued)
  • Page 60 ThePoint™ Series Control Drawings Adding a Padded Capacitor (Continued)
  • Page 61: Dual Seal Assembly For 700 Series Sensing Elements

    ThePoint™ Series Control Drawings Dual Seal Assembly for 700 Series Sensing Elements...
  • Page 62 ThePoint™ Series Control Drawings Dual Seal Assembly (Continued)
  • Page 63 ThePoint™ Series Control Drawings Dual Seal Assembly (Continued)
  • Page 64 Appendix A...
  • Page 65: Appendix A Shortening Or Lengthening Sensing Element

    ThePoint™ Series Appendix A Shortening or Lengthening Sensing Element CAUTION: The length of either Flush Sensing Elements or Insulated Sensing Elements can Not be changed. Cable Sensing Elements can only be shortened. Instructions are included with each unit. The Need Sometimes your application calls for probe lengths other than the standard 18-inch or longer insertion lengths supplied.
  • Page 66: Issue

    AMETEK, Inc. TERMS AND CONDITIONS OF SALE THE FOLLOWING TERMS/CONDITIONS, TOGETHER WITH ANY OTHER TERMS/CONDITIONS SPECIFICALLY AGREED TO IN WRITING BY SELLER, SHALL APPLY TO ALL ORDERS (“Order(s)”) FROM, AND SALES OF PRODUCTS (“Products”) OR SERVICES (“Services”) TO BUYER. ANY ACCEPTANCE OF ANY ORDER OF BUYER IS CONDITIONED UPON THESE TERMS/CONDITIONS. ANY ADDITIONAL OR DIFFERENT TERMS/CONDITIONS PROPOSED BY BUYER IN ANY DOCUMENT ARE OBJECTED TO AND SHALL NOT BE BINDING UPON SELLER.
  • Page 67 WHETHER ORAL, WRITTEN, EXPRESS, IMPLIED OR STATUTORY. Seller’s prior written consent. Upon completion of Order, Buyer shall promptly return all Data to Seller together with all copies or reprints thereof NO IMPLIED OR STATUTORY WARRANTIES OF then in Buyer's possession or control, and Buyer shall thereafter make no MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE future use, either directly or indirectly, of any Data or any information SHALL APPLY.
  • Page 68 Telephone: +1 215-674-1234 Fax: +1 215-674-2731 An ISO 9001 Certified Company E-mail: drexelbrook.info@ametek.com Website: www.drexelbrook.com...

Table of Contents