1)
Setup the test area according to IEC 61000-4-2 or ISO 10605 and the test plan
2)
Ensure cable and base station earth are connected
3)
Connect the proper Air (round) or Contact (pointed) discharge tip to the front of the discharge pistol
4)
Insert the proper R/C network according to the standard and test plan.
5)
Connect the supplied power supply
6)
Press POWER/POWER ON switch.
7)
Press the TEST ON or INTERLOCK button and release the "Emergency Power Off" switch if equipped
8)
Wait for the Self Check and Calibration to finish.
9)
On the display, tap on the voltage to set the voltage
10) Press the trigger to fire a pulse
Quick Start and Safety Guide
Figure 2 - User Interface, 30 kV models