Refill Indicator; Dispensing Liquid Cleaning Agent - Miele PG 8504 Operating Instructions Manual

Commercial dishwasher for laboratory glassware and utensils
Table of Contents

Advertisement

Adding and dispensing chemical agents
Checking
consumption

Refill indicator

Dispensing liquid
cleaning agent
52
Push the suction lance into the opening of the container and secure
the lid. Observe the colour coding.
Feed the suction lance into the container until it reaches the
bottom.
Wipe up any spilled process chemical thoroughly.
Place the container on the floor next to the machine or in an
adjacent cupboard. The container must not be placed on top of or
above the machine. Make sure that the dispensing hose is not
kinked or trapped.
The dispensing system must then be vented (see "Settings /DOS
venting").
Check consumption regularly by checking the fill levels in the supply
containers and replace containers in good time to avoid the
dispensing system being sucked completely dry.
When the fill level is low in the
cleaning agent you are reminded to replenish it.
Confirm the message shown with OK and
replenish the liquid cleaning agent as described.
If the liquid cleaning agent has run out, the machine will be locked
for further use.
It will be ready for use again when the supply container has been
replaced.
For adjusting the dispensing concentration, see "Further settings/
Additional functions/Dispensing systems".
DOS 1
supply container for liquid
DOS
Refill