COOK Medical Duette DT-6-5F Quick Reference Manual page 2

Multi-band mucosectomy

Advertisement

Prep 6
Place the trigger cord into the slot on the spool
of the ligator handle and pull down until the
knot is seated in the hole of the slot. The knot
must be seated into the hole for the handle to
function properly.
Snare - Prep 1
Fully retract and extend the snare to confirm
smooth operation.
Customer Service
EMEA:
EDI – www.cookmedical.com/edi.do
Distributors: +353 61239240, ssc.distributors@cookmedical.com
Austria: +43 179567121, oe.orders@cookmedical.com
Belgium: +32 27001633, be.orders@cookmedical.com
Denmark: +45 38487607, da.orders@cookmedical.com
Finland: +358 972519996, fi.orders@cookmedical.com
France: +33 171230269, fr.orders@cookmedical.com
Germany: +49 6950072804, de.orders@cookmedical.com
Hungary: +36 17779199, hu.orders@cookmedical.com
Ireland: +353 61239252, ie.orders@cookmedical.com
Italy: +39 0269682853, it.orders@cookmedical.com
Netherlands: +31 202013367, nl.orders@cookmedical.com
Norway: +47 23162968, no.orders@cookmedical.com
Spain: +34 912702691, es.orders@cookmedical.com
Sweden: +46 858769468, se.orders@cookmedical.com
Switzerland – French: +41 448009609, fr.orders@cookmedical.com
Switzerland – Italian: +41 448009609, it.orders@cookmedical.com
Switzerland – German: +41 448009609, de.orders@cookmedical.com
United Kingdom: +44 2073654183, uk.orders@cookmedical.com
M E D I C A L
cookmedical.com
Prep 7
With the handle in the two-way position, slowly
rotate the handle clockwise to wind the trigger
cord onto the handle spool. Note: Care must be
exercised to avoid deploying a band while
winding the trigger cord.
Snare - Prep 2
Slide the adjustable marker located in the
handle shaft to establish a reference point
indicating full retraction of the snare into the
sheath. The grid on the handle may be used
to set up reference points, establishing the
thickness of the tissue being excised during
the procedure. To utilize these markings, fully
advance and retract the snare to establish the
markings which indicate snare position.
Americas:
EDI – www.cookmedical.com/edi.do
Phone: +1 812.339.2235, 800.457.4500, Fax: 800.554.8335
E-mail: orders@cookmedical.com
Australia:
Phone: +61 734346000, 1800777222, Fax: +61 734346001, 1800077283
E-mail: cau.custserv@cookmedical.com
Prep 8
Check the endoscopic view. To maximize
visualization, the position of the trigger cord
may be altered by rotating the barrel.
Refer to current instructions
for detailed system use.