Warranty

#
Problem
Display reaction is erratic:
"0.0'' or "OVER'' appear on
7
the screen before the
apical constriction is
reached.
11
Warranty
Propex Pixi
The accessories (cables, etc.) are warranted for 6 months from the date
of purchase.
The warranty is valid for normal usage conditions. Any modification or
accidental damage will void the warranty.
12

Disposal of the Product

Recycling: PLEASE DO NOT THROW AWAY! This product and all its
components must be recycled through your supplier.
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
Possible cause
Short circuit due to
excess liquid
(irrigation solution,
saliva, blood) in the
pulp chamber.
A direct contact of
the file with the
gingiva or gingival
proliferations, e.g. a
fractured metal
crown.
A direct contact of
the file with metal
restorations (crown,
parapulpal post,
amalgam filling).
®
is warranted for 12 months from the date of purchase.
Solution
Dry the access cavity with
a cotton pellet / air-blower.
In case of excess bleeding
wait until it has stops.
For isolation:
adequate preparation
filling of access cavity;
use a rubber dam.
Isolate the file by inserting
it into a small polyvinyl tube
before use.