Parameter assignment
8.3 Parameter assignment over device with display
8.3.13

One-point calibration input 1 [14]

Calibrates the lower calibration point at input 1. The device moves the characteristic by the
difference between the original and the new calibration point. The result of the one-point
calibration is saved in the TRIM INPUT 1 parameter.
Indication on the dis‐
play:
The two-point calibration via remote operation calibrates the low and high calibration point.
See also
Setting two-point calibration (Page 100)
Setting one-point calibration (Page 99)
8.3.14

One-point calibration input 2 [15]

Calibrates the lower calibration point at input 2. The device moves the characteristic by the
difference between the original and the new calibration point. The result of the one-point
calibration is saved in the TRIM INPUT 2 parameter.
Indication on the dis‐
play:
The two-point calibration via remote operation calibrates the low and high calibration point.
See also
Setting two-point calibration (Page 100)
Setting one-point calibration (Page 99)
8.3.15

Change User PIN [16]

8.3.15.1

Change user PIN

Used to change the User PIN.
Setting range:
Factory setting:
Requirement
The "User PIN (Page 91)" parameter is enabled.
88
TRIM INPUT 1
TRIM INPUT 2
1 to 65535
2457
SITRANS TS100/TS200/TS300/TS500/TSinsert/TSthermowell
Operating Instructions, 08/2020, A5E47810090-AA